IM 12B07D02-01E
Appendix 10-3
10-2. Configuration checklist for PH202G
Standard Configuration
Options
Reference for
change
Measured Variable(s)
primary inputs
pH, ORP and Temp
pH range
0-14 pH
any span within -2-16 pH
“output”
pH range linearized
disabled
21 point table
codes 31& 35
ORP range
-500 to 500 mV
spans up to 3000 mV between
“output”
-1500 to 1500mV
Temperature range
-30-140°C
Temperature unit
Celsius
Fahrenheit
code 11
mA Outputs
analog output
4- 20 mA for pH
pH/ORP/(parameter 2)
code 01, 02, 31
output linearization
disabled
pH/ORP
codes 35
Communication
digital interface
disabled
HART
60
communication software
external
HHC or PC202
contact factory
variables on display
pH/ORP and temp
pH/ORP, parameter 2, mA output
“display”
SL, AP, Z1, Z2 etc.
burn out
disabled
burn low (3.6 or 3.9)/ high (21) on mA output code 32
password protection
disabled
for maint/ comm./ serv level
code 52
autoreturn
return to measure in 10 min. enable or disable
code 50
add. functions in MAINT
disabled
Impedance check start
code 51
Diagnostics
impedance checking
active
enable or disable
code 03 & 04
check on calibration data
active
enable or disable
code 05
check on stability
0.02 pH per 5 s
choose stability level
code 20
display calibration log.
enabled with logbook
diagnostics setup
codes 03, 04 & 05
Compatibility
pH or ORP
glass sensor/metal electrode pH or ORP
code 01
temperature sensor
Pt 1000
Ω
Pt1000; Pt100, etc
code 10
other sensors
enamel sensors (Pfaudler)
ITP & impedance check setup
codes 23, 03 & 04
2nd parameter
disabled
pH & ORP/ pH &rH
code 02
manual temp. comp.
disabled
disable or enable
“temp”
Special Features
buffer table configuration
NIST standard
fully configurable
codes 24, 25 & 26
temperature calibration
none
adju/- 20 °C
code 12
zero point calibration
disabled
disable or enable
code 27
call for maintenance
set time interval 1 - 250 days
code 55
HOLD during maintenance
hold last or hold fix
“hold”
process temp. compensation disabled
set temperature coefficient
code 13
logbook
disabled
2 x 50 events
code 61, 62
mA
mA
mA
mA
Содержание PH202G (S)
Страница 5: ...IM 12B07D02 01E ...
Страница 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Страница 13: ...IM 12B07D02 01E 1 4 Introduction ...
Страница 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Страница 96: ...IM 12B07D02 01E 10 10 Appendix ...
Страница 116: ...IM 12B07D02 01E 12 4 Appendix ...
Страница 120: ...IM 12B07D02 01E 12 8 Appendix ...
Страница 124: ...IM 12B07D02 01E 12 12 Appendix ...
Страница 128: ...IM 12B07D02 01E 12 16 Appendix ...