IM 12B07D02-01E
Appendix 10-1
10. APPENDIX
10-1. User setting table
FUNCTION
SETTING DEFAULTS
USER SETTINGS
Parameter specific functions
01 *PH.ORP 0
pH
02 *PRM2 0
Off
03 *Z1.CHK
1.1.1
High range, TC on
check
on,
04 *Z2.CHK
0.0.1
Low range, TC off
check off no TC
05 *CAL.CK
1.1
AP on, Slope on
Temperature functions
10 *T.SENS 0
Pt1000
11 *T.UNIT 0
°C
12 *T.ADJ
None
13 *T.COMP 0
Off
*T.COEF
-0.00 pH/10°C
Calibration functions
20 *
∆
t.SEC 5
Sec
*
∆
PH 0.02 pH
21 *AS.LOW -120
mV
*AS.HI 120
mV
22 *SL.LOW 70
%
*SL.HI 110
%
23 *ITP
7.00
pH
*SLOPE 100.0 %
*ASP.1D
0.0
mV
*ASP.mV
mV
24 *BUF.ID 4
NIST
4
25 *BUF.ID 7
NIST
7
26 *BUF.ID 9
NIST
9
27 *ZERO.P 0
disabled
mA outputs
31 *OUTP.F 0
pH
(ORP)
32 *BURN 0
off
35 *TABLE
21 pt table see code 31
mA
mA
mA
mA
mA
mA