background image

7/12

IM MT300-71EN

Explanation

Wiring Method and Instrument Configuration

Adjusting the D/A Output

•  Connect this instrument’s D/A output terminal to the 3458A digital multimeter’s voltage input terminal.

   

DMM (3458A)

H

L

MT300

D/Aout

+ -

•  Set the 3458A measurement range to 10 VDC range and the integration time to 50NPLC.

Adjusting the Voltage Measurement

•  Connect this instrument’s voltage input terminal, the 3458A digital multimeter’s voltage input terminal, and 

the GS200 standard DC voltage current source’s output terminal.

   

SOURCE (GS200)

H

L

MT300

V

DMM (3458A)

H

L

A COM

•  Set the 3458A measurement range to 10 VDC range and the integration time to 50NPLC.

Adjusting the Current Measurement

•  Connect this instrument’s current input terminal, the 3458A digital multimeter’s voltage input terminal, the 

GS200 standard DC voltage current source’s output terminal, and the 2792A06 standard resistor.

   

SOURCE (GS200)

H

L

MT300

V

DMM (3458A)

H

L

A COM

100Ω

•  Set the 3458A measurement range to 10 VDC range and the integration time to 50NPLC.

Содержание MT300

Страница 1: ...The following options are installed in the MT300 D A output adjustment DA option Voltage current measurement adjustment DM option Equipment Used Digital multimeter 3458A Keysight or equivalent Standa...

Страница 2: ...ess the Range soft key to select the 2 V range Output value 4 Press the Zero soft key The zero point 0 V is output and the input screen appears 5 When the digital multimeter display stabilizes use the...

Страница 3: ...soft key The voltage current measurement adjustment menu appears Note If the warning code 59 The measured value is out of range appears when you set an adjustment value the value may be exceeding the...

Страница 4: ...A range 13 Connect this instrument digital multimeter standard DC voltage current source and standard resistor For the wiring method and instrument setup see Explanation 14 Press the PlusFull soft key...

Страница 5: ...and 20 mA ranges are saved and the screen returns with adjustment menu Ending Adjustment Mode 1 On the adjustment menu press ESC 2 Press the OK soft key The screen returns the top menu Canceling the A...

Страница 6: ...n the Next soft key The following menu appears 3 Press the Maintenance soft key The Maintenance menu appears 4 Press the Calibration Date soft key The following menu appears 5 If you calibrated the D...

Страница 7: ...digital multimeter s voltage input terminal and the GS200 standard DC voltage current source s output terminal SOURCE GS200 H L MT300 V DMM 3458A H L A COM Set the 3458A measurement range to 10 VDC ra...

Страница 8: ...Sets the full scale negative adjustment value of the voltage current measurement 10 CALibrate DMM ADJust PFULl Sets the full scale positive adjustment value of the voltage current measurement 10 CALi...

Страница 9: ...te DA POINt command to set the adjustment point to PFULL If the adjustment point is not PFULL a 221 Setting conflict error will occur CALibrate DA ADJust ZERO Function Sets the zero point adjustment v...

Страница 10: ...t target to DA If the adjustment target is not DA a 221 Setting conflict error will occur When executed the adjustment target is set to NONE CALibrate DMM ADJust MFULl Function Sets the full scale neg...

Страница 11: ...ALibrate DMM POINt PFULL MFULL CALibrate DMM POINt PFULL Full scale positive value MFULL Full scale negative value Example CALIBRATE DMM POINT PFULL CALIBRATE DMM POINT CALIBRATE DMM POINT PFULL Descr...

Страница 12: ...sure to write the adjustment values using the appropriate command before changing the adjustment target On models without the DM option sending a CALibrate FUNCtion DMM command will cause a 141 Inval...

Отзывы: