:CALibrate:DA:POINt
Function
Sets or queries the adjustment point of the D/A
output range to be adjusted.
Syntax
:CALibrate:DA:POINt {PFULL|MFULL
|ZERO}
CALibrate:DA:POINt?
PFULL: Full-scale positive value
MFULL: Full-scale negative value
ZERO: Zero point
Example
:CALIBRATE:DA:POINT PFULL
:CALIBRATE:DA:POINT?
-> :CALIBRATE:DA:POINT PFULL
Description • On models without the /DA option, a “241:
Hardware missing” error will occur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DA. If
the adjustment target is not DA, a “221: Setting
conflict” error will occur. A query returns the
current value.
:CALibrate:DA:RANGe
Function
Sets or queries the D/A output range to be
adjusted.
Syntax
:CALibrate:DA:RANGe <
Voltage
>
:CALibrate:DA:RANGe?
<Voltage> = 2, 5 (V)
Example
:CALIBRATE:DA:RANGE 5V
:CALIBRATE:DA:RANGE?
-> :CALIBRATE:DA:RANGE 5.0E+00
Description • On models without the /DA option, a “241:
Hardware missing” error will occur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DA. If
the adjustment target is not DA, a “221: Setting
conflict” error will occur. A query returns the
current value.
:CALibrate:DA:WRITe
Function
Writes the adjustment value of D/A output.
Syntax
:CALibrate:DA:WRITe
Example
:CALIBRATE:DA:WRITE
Description • On models without the /DA option, a “241:
Hardware missing” error will occur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DA. If
the adjustment target is not DA, a “221: Setting
conflict” error will occur.
• When executed, the adjustment target is set to
NONE.
:CALibrate:DMM:ADJust:MFULl
Function
Sets the full-scale negative adjustment value of
the voltage/current measurement.
Syntax
:CALibrate:DMM:ADJust:MFULl <NRf>
<NRf> =
5 V range
−4.00000 (V) to −6.00000 (V) (0.00001 V setting
resolution)
20 mA range
−16.0000 (mA) to −24.0000 (mA) (0.00001 mA
setting resolution)
Example
:CALIBRATE:DMM:ADJUST:MFULL −5.00000
Description • On models without the /DM option, a “241:
Hardware missing” error will occur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DMM.
If the adjustment target is not DMM, a “221:
Setting conflict” error will occur.
• Use the
:CALibrate:DMM:POINt
command
to set the adjustment point to MFULL. If the
adjustment point is not MFULL, a “221: Setting
conflict” error will occur.
:CALibrate:DMM:ADJust:PFULl
Function
Sets the full-scale positive adjustment value of
the voltage/current measurement.
Syntax
:CALibrate:DMM:ADJust:PFULl <NRf>
<NRf> =
5 V range
4.00000 (V) to 6.00000 (V) (0.00001 V setting
resolution)
20 mA range
16.0000 (mA) to 24.0000 (mA) (0.00001 mA
setting resolution)
Example
:CALibrate:DMM:ADJust:PFULl 5.00000
Description • On models without the /DM option, a “241:
Hardware missing” error will occur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DMM.
If the adjustment target is not DMM, a “221:
Setting conflict” error will occur.
• Use the
:CALibrate:DMM:POINt
command
to set the adjustment point to PFULL. If the
adjustment point is not PFULL, a “221: Setting
conflict” error will occur.