3/12
IM MT300-71EN
10.
Press the
MinusFull
soft key. The full-scale negative value (−2 V) is output, and the input screen
appears.
11.
When the digital multimeter display stabilizes, use the
arrow
(▲, ▼, ◄, and ►) keys to input the value
displayed on the digital multimeter.
12.
Press
ENTER
. The full-scale negative adjustment value is set.
13.
Press the
Range
soft key to select the 5 V range.
14.
Repeat steps 4 to 12, set the adjustment values (zero point, full-scale positive and negative values) of
the 5 V range.
15.
Press the
OK
soft key. The adjustment values of the 2 V and 5 V ranges are saved, and the screen
returns with adjustment menu.
Adjusting the Voltage/Current Measurement
1.
Press the
DMM
soft key. The voltage/current measurement adjustment menu appears.
Note
If the warning code “59: The measured value is out of range” appears when you set an adjustment value, the value may
be exceeding the adjustment range. Use the:
CALibrate:DMM:INITialize
command to initialize the voltage/current
measurement adjustment values. When you initialize the adjustment values, they are returned to their factory default
conditions. Adjust the voltage/current measurement again.
Adjusting the Voltage Measurement
2.
Press the
Range
soft key to select the 5 V range.
Measurement value
When you enter adjustment
values, they are reflected in
the measurement value.
3.
Connect this instrument, digital multimeter, and standard DC voltage current source.
For the wiring method and instrument setup, see “Explanation.”
4.
Press the
PlusFull
soft key.
5.
On the standard DC voltage current source, 5 V.