6. LIST OF PARAMETERS
Parameter Display
Item
MODEL
Model
TAG NO
Tag no.
SELF CHK
Self-check result
A
DISPLAY1
Display1
A01 INPUT1
Input value 1
A35 COUNTER1
Integrating counter 1
(*3)
A36 COUNTER2
Integrating counter 2
(*3)
A37 COUNTER3
Integrating counter 3
(*3)
A38 COUNTER4
Integrating counter 4
(*3)
A54 STATUS
Status
(*1)
A56 REV NO
Rev. no.
A58 MENU REV
MENU REV
A60
SELF CHK
Self-check result
B
DISPLAY2
Display2
B01
INPUT1
Input value 1
B35
COUNTER1
Integrating counter 1
(*3)
B36
COUNTER2
Integrating counter 2
(*3)
B37
COUNTER3
Integrating counter 3
(*3)
B38
COUNTER4
Integrating counter 4
(*3)
B60 SELF CHK
Self-check result
D
SET (I/O)
Setting (I/O)
(*2)
D01 TAG NO.1
Tag no. 1
D02 TAG NO.2
Tag no. 2
D03 COMMENT1
Comment 1
D04 COMMENT2
Comment 2
D20 UNIT
Input display unit
D50 OUT1 DR
Pulse logic (D/R)
D52
PULSE RATE
Pulse rate
D53
PULSE TYPE
Pulse width type
D54
PULSE WIDTH
Pulse width time
D58
INPUT FILTER
Input filter
D60
SELF CHK
Self-check result
F
SET(COM)
Setting (communication)
(*2)
F01
PROTOCOL
Communication protocol
F02 ADDRESS
Address
F03
BAUD RATE
Baud rate
F04 PARITY
Parity
F05
DATA LEN
Data Length
F06
STOP BIT
Stop bit
F07 INPUT DEC PT
Decimal point position of input
F60
SELF CHK
Self-check result
P
ADJUST
Adjustment
P60
SELF CHK
Self-check result
*1 The Status is displayed for service personnel to see history
records.
*2
There are items not displayed depending on what output-2 is
specified.
*3
The 5-byte hexadecimal integrating counter value is displayed in
decimal format in integrating counters 1 through 4.
7. MAINTENANCE
The product starts running immediately when the power is turned
on; however, it needs 10 to 15 minutes of warm-up before it
meets the specified performance.
7.1 Calibration Apparatus
• A calibrator (YOKOGAWA CA500 or equivalent)
7.2 Calibration Procedure
1.
Connect the instruments as shown below. First adjust the
output-1 signal and then the output-2 signal.
10
11
3 2 1
4
5
6
7
8
9
Power supply
11
10
8
L
+
N
–
GND
3
4
+
–
Input
Calibrator
Calibrator
2
5
Output-2
+
–
Calibrator
7
9
Output-1
+
–
2. Produce a rectangular pulse of any frequency from the
calibrator to measure the value using a calibrator, then check
that the frequency (input frequency x set rate) is output.
8. SAFETY AND EMC STANDARDS
The following will be acquired.
Safety:
IEC/EN 61010-1 compliance (CE)
IEC/EN 61010-2-030 compliance (CE)
CAN/CSA-C22.2 No.61010-1 (CSA)
CAN/CSA-C22.2 No.61010-2-030 (CSA)
UL61010-1 (CSA NRTL/C)
UL61010-2-030 (CSA NRTL/C)
Overvoltage category: II
Pollution degree: 2
Measurement category: O (other)
Rated power supply voltage: 15-30VDC( ± 10% ) 3.7W
100-240VAC/DC( ± 10% )
50/60 Hz 3.7 W 9.2 VA
Rated voltage (Input signal) : 50 V DC
Rated voltage (Communication output signal) : 5 V DC
Rated voltage (
Voltage Pulse
output signal) : 26.4 V DC
Rated transient overvoltage: 1500 V (*)
* This is a reference safety standard value for the
measurement Category O. This value is not necessarily
a guarantee of instrument performance.
CAUTION
This instrument is for Measurement Category O
(other). Do not use it for measurements in locations
falling under Measurement Categories II, III, and IV.