6.  LIST OF PARAMETERS

Parameter Display

Item

MODEL

Model

TAG NO

Tag no.

SELF CHK

Self-check result

A

DISPLAY1

Display1

A01 INPUT1

Input value 1

A35 COUNTER1

Integrating counter 1

(*3)

A36 COUNTER2

Integrating counter 2

(*3)

A37 COUNTER3

Integrating counter 3

(*3)

A38 COUNTER4

Integrating counter 4

(*3)

A54 STATUS

Status

(*1)

A56 REV NO

Rev. no.

A58 MENU REV

MENU REV

A60

SELF CHK

Self-check result

B

DISPLAY2

Display2

B01

INPUT1

Input value 1

B35

COUNTER1

Integrating counter 1

(*3)

B36

COUNTER2

Integrating counter 2

(*3)

B37

COUNTER3

Integrating counter 3

(*3)

B38

COUNTER4

Integrating counter 4

(*3)

B60 SELF CHK

Self-check result

D

SET (I/O)

Setting (I/O)

(*2)

D01 TAG NO.1

Tag no. 1

D02 TAG NO.2

Tag no. 2

D03 COMMENT1

Comment 1

D04 COMMENT2

Comment 2

D20 UNIT

Input display unit

D50 OUT1 DR

Pulse logic (D/R)

D52

PULSE RATE

Pulse rate

D53

PULSE TYPE

Pulse width type

D54

PULSE WIDTH

Pulse width time

D58

INPUT FILTER

Input filter

D60

SELF CHK

Self-check result

F

SET(COM)

Setting (communication)

(*2)

F01

PROTOCOL

Communication protocol

F02 ADDRESS

Address

F03

BAUD RATE

Baud rate

F04 PARITY

Parity

F05

DATA LEN

Data Length

F06

STOP BIT

Stop bit

F07 INPUT DEC PT

Decimal point position of input

F60

SELF CHK

Self-check result

P

ADJUST

Adjustment

P60

SELF CHK

Self-check result

*1 The Status is displayed for service personnel to see history 

records.

*2

There are items not displayed depending on what output-2 is 

specified.

*3

The 5-byte hexadecimal integrating counter value is displayed in 

decimal format in integrating counters 1 through 4.

7.  MAINTENANCE

The product starts running immediately when the power is turned 

on; however, it needs 10 to 15 minutes of warm-up before it 

meets the specified performance.

7.1  Calibration Apparatus

 • A calibrator (YOKOGAWA CA500 or equivalent)

7.2  Calibration Procedure

1. 

Connect the instruments as shown below. First adjust the 

output-1 signal and then the output-2 signal.

  

10

11

3 2 1

4

5

6

7

8

9

Power supply

11

10

8

L

+

N

GND

3

4

+

Input

Calibrator

Calibrator

2

5

Output-2

+

Calibrator

7

9

Output-1

+

2.  Produce a rectangular pulse of any frequency from the 

calibrator to measure the value using a calibrator, then check 

that the frequency (input frequency x set rate) is output.

8.  SAFETY AND EMC STANDARDS

The following will be acquired.

Safety:

IEC/EN 61010-1 compliance (CE)

IEC/EN 61010-2-030 compliance (CE)

CAN/CSA-C22.2 No.61010-1 (CSA)

CAN/CSA-C22.2 No.61010-2-030 (CSA)

UL61010-1 (CSA NRTL/C)

UL61010-2-030 (CSA NRTL/C)

Overvoltage category: II

Pollution degree: 2

Measurement category: O (other)

Rated power supply voltage:  15-30VDC( ± 10% ) 3.7W 

100-240VAC/DC( ± 10% ) 

50/60 Hz 3.7 W 9.2 VA 

Rated voltage (Input signal) : 50 V DC

Rated voltage (Communication output signal) : 5 V DC

Rated voltage (

Voltage Pulse

 output signal) : 26.4 V DC

Rated transient overvoltage: 1500 V (*)

* This is a reference safety standard value for the 

measurement Category O. This value is not necessarily 

a guarantee of instrument performance.

 CAUTION

This instrument is for Measurement Category O 

(other). Do not use it for measurements in locations 

falling under Measurement Categories II, III, and IV.

    

Internal Wiring

Outlet

Entrance 

Cable

III

T

O

II

IV