CAUTIONS WHEN USING EXPLOSION-PROTECTED INSTRUMENTS
The GC8000 is designed to protect against explosion.
When the analyzer is used in a hazardous area, observe the following precautions.
Since the applicable standard differs depending on the specifications of the analyzer to be used,
check the specifications of your analyzer.
l
Type of explosion protection
To assure explosion protection, the GC8000 has a pressurized and flameproof enclosure,
meeting the following standards (flameproof
enclosure is not provided in FM-Y, CSA-Y):
<GC8000-F (FM-X), GC8000-G (FM-Y)>
Type X pressurization and Explosionproof for Class I, Division 1, Group B, C and D
(Described as FM-X hereafter)
Type X and Y pressurization for Class I, Division 1, Group B, C & D
(Described
as
FM-Y
hereafter)
T1 (programmed-temperature oven 320°C max., isothermal oven 225°C max., liquid-sample
valve 250°C max.)
T2 (programmed-temperature oven 225°C max., isothermal oven 225°C max., liquid-sample
valve 225°C max.)
T3 (programmed-temperature oven 145°C max., isothermal oven 145°C max., liquid-sample
valve 145°C max.)
T4 (programmed-temperature oven 95°C max., isothermal oven 95°C max., liquid-sample
valve 95°C max.)
<GC8000-C (CSA-X), GC8000-D (CSA-Y)>
Type X pressurization and Explosionproof for Class I, Division 1, Group B, C and D
(Described as CSA-X hereafter)
Type X and Y pressurization for Class I, Division 1, Group B, C & D
(Described
as
CSA-Y
hereafter)
T1 (programmed-temperature oven 320°C max., isothermal oven 225°C max., liquid-sample
valve 250°C max.)
T2 (programmed-temperature oven 225°C max., isothermal oven 225°C max., liquid-sample
valve 225°C max.)
T3 (programmed-temperature oven 145°C max., isothermal oven 145°C max., liquid-sample
valve 145°C max.)
T4 (programmed-temperature oven 95°C max., isothermal oven 95°C max., liquid-sample
valve 95°C max.)
<GC8000-A (ATEX)>
ATEX: Group II Category 2G, DEKRA 11ATEX0238 X *
Ex db pxb II B +H
2
T1 Gb (programmed-temperature oven 320°C max., isothermal oven
225°C max., liquid-sample valve 250°C max.)
Ex db pxb II B + H
2
T2 Gb (programmed-temperature oven 225°C max., isothermal oven
225°C max., liquid-sample valve 225°C max.)
Ex db pxb II B + H
2
T3 Gb (programmed-temperature oven 145°C max., isothermal oven
145°C max., liquid-sample valve 145°C max.)