Setting the Time Scale
Set the time per division of the trend waveform to 2 minutes.
The sampling interval (the time
corresponding to 1 dot) is 4 s when the trend
interval is 2 min.
30 dots × 4 s = 2 min
30 dots/div
Setup Item
Description
Number in the Figure
Trend interval
Set the time per division to 2 minutes.
The waveform is updated at every 4 s.
1
(1) Trend interval
Press
MENU
(to switch to setting mode) > select the
Menu
tab >
Display
>
Trend/
Save interval
.
1
Operation complete.
For more details of the setting trand interval, see Section 6.1 “Setting the Recording
Conditions of the Measured Data”, in the
FX1000 User’s Manual
, IM 04L21B01-01EN.
If you start memory sampling after configuring the settings as shown above, the
measured values are displayed as a waveform (trend display) and are recorded to the
internal memory. The data recorded to internal memory is separated into files. Each
file contains one hour's worth of data. To change the file size, see Section 6.1 “Setting
the Recording Conditions of the Measured Data”, in the
FX1000 User's Manual
, IM
04L21B01-01EN.
On FXs that have a CF card slot, if a CF card has been inserted in the slot, files are
saved automatically to the "DATA0" folder on the CF card. To change how data is saved
to the CF card, see Section 6.2 “Setting the Method for Saving the Data”, in the
FX1000
User's Manual
, IM04L21B01-01EN.
If you want to save setup data on an FX that has a CF card slot, follow the procedure
under "Saving the Setup Data" on the next page.