<4. Setting converter>
4-19
IM 12A01G01-03EN
1st Edition : Mar. 25, 2021-00
Figure 4.13
Address CH assignment and mA output
n
MODBUS - Converter
FLXA402T as slave communicates with HOST devices, via MODBUS RS-485, or Ethernet
(Modbus TCP/IP). Select a network system according to the hardware architecture you use.
For further technical information on MODBUS, read
TI 12A01F01-62EN.
4.6.2
Ethernet setting (E)
Modbus TCP/IP is available by Ethernet cable-connection between the converter and master
devices.
Interface
Ethernet
Maximum session times
2
Protocol
Modbus//TCP
Port number
502
Ethernet 10/100 Mbps speed rate, Internet Protocol version 4 (IPv4)
Set “DHCP enable” to “On” if you want DHCP server to assign IP address automatically. Set it
“Off” (Off is default) if you use a static IP address.
When you use a static IP address, set “Subnet mask” and “Default gateway”, according to
Ethernet you use.
The next table shows default of Ethernet parameter.
Parameter name
Default
DHCP
Off
IP address
192.168.1.10
Subnet mask
255.255.255.0
Default gateway
192.168.1.1
NOTE
After rebooting FLXA402T (Power OFF/ON), Ethernet setting is valid.