50
4.
BASIC OPERATION
IM FG410-01EN
4.4.8 Setting DC offset
a) Setting method
1.
Press the
OFFSET shortcut key to
open the DC offset input field.
Alternatively, select the [Offset] field and
then press the
ENTER key to open
the DC offset input field.
Even if there are several setting screens,
the [Offset] field is always displayed on the
first page.
If [Low] is displayed in the [Offset] field
and the low level instead of the offset is
displayed, press the
OFFSET key
once more to switch to the DC offset
display.
2.
Select the digit to be changed with the
left/right arrow keys, and increment/
decrement the value with the
up/down arrow keys or the
modify knob. The
change is instantly reflected to the output.
Alternatively, enter a numeric value using the
...
numeric keypad. Then press the
ENTER key or one of the unit keys (soft keys) to set the input value and reflect it to
the output.
b) AC+DC limitation
The maximum value for [AC amp DC offset] is limited to ±10 V/open.
For example, if the AC amplitude is 5 Vp-p/open, the DC offset range is limited to the range
from
−
7.5 V/open to +7.5 V/open.
The maximum value varies according to the output voltage range setting and the external
addition input setting.
DC offset input
field
OFFSET
ENTER
OFFSET
0
9
ENTER
Содержание FG400
Страница 1: ...User s Manual FG400 Arbitrary Function Generator User s Manual Basic IM FG410 01EN 1st Edition ...
Страница 8: ...VI IM FG410 01EN ...
Страница 159: ...7 17 General Characteristics 147 IM FG410 01EN External dimensions FG410 Unit mm Rear view ...
Страница 160: ...148 7 SPECIFICATIONS IM FG410 01EN External dimensions FG420 Unit mm Rear view ...
Страница 161: ...RACK MOUNTING DIMENSIONS 149 IM FG410 01EN Inch rack mounting dimensions for 1 unit Unit mm RACK MOUNTING DIMENSIONS ...
Страница 163: ...RACK MOUNTING DIMENSIONS 151 IM FG410 01EN Millimeter rack mounting dimensions for 1 unit Unit mm ...
Страница 169: ...157 IM FG410 01EN ...
Страница 171: ......