<4. OPERATION>
4-3
IM 12F5A2-01E
NOTE
For the FC500G-1, measuring ranges of 0 to 5 and 0 to 10 mg/l cannot be used. For the
FC500G-5, measuring ranges of 0 to 1, 0 to 2, and 0 to 3 mg/l cannot be used.
4.1.4
Setting of “Applied Voltage Compensation/Applied
Voltage Fixation’’ Selection Jumper
The applied voltage can be fixed to a definite value without applied voltage compensation. When
the applied voltage compensation/applied voltage fixation transfer jumper shown in Figure 4.5 is
set to the side B (lower), the applied voltage is fixed to the starting voltage value.
Applied voltage
compensation
Side A
J5
Applied voltage
fixation
Side B
J5
J5
F4-5.ai
Figure 4.5
“Applied Voltage Compensation/Applied Voltage Fixation” Selection Jumper
NOTE
If conductivity exceeds 300 μS/cm, applied voltage should be changed to fixation.