21
IM 12D06D05-01E
5
MENU
STRUCTURE
COMMISSIONING
Menu
Parameter Default
Range
values
min.
max.
Configure Sensor
Cell constant
1.88 cm
-1
0.2 cm
-1
50.0 cm
-1
Temp. Comp.
Reference Temp.
25ºC, 77ºF
0ºC, 32ºF
100ºC, 211ºF
Manual Comp.
Manual Temp.
25ºC, 77ºF
-20ºC, -3.9ºF
139ºC, 284ºF
Temp. Coef
T.C.methods 1
2.10%/ºC
0%/ºC, 0%/ºF
3 . 5 % / º C , 2 . 0 % / º F
Temp. Coef
T.C.methods 2
2.10%/ºC
0%/ºC, 0%/ºF
3.5%/ºC, 2.0%/ºF
Enter
Measurement setup
Temperature settings
Temp. Compensation
Calibration settings
Concentration
Measure Conductivity only