3-4. Wiring the contact signals
3-4-1. General precautions
The contact output signals consist of volt-
age-free relay contacts for switching electrical
appliances (SPDT). They can also be used as
digital outputs to signal processing equipment
(such as a controller or PLC). It is possible to
use multi-core cables for the contact in and
output signals and shielded multi-core cable for
the analog signals.
3-4-2. Contact outputs
The EXAxt 450 unit’s four contacts (switches)
that can be wired and configured to suit user
requirements. Contact S4 is programmed as
a fail-safe contact. Please refer to section 5-7,
Contact output setup for functionality descrip-
tion.
Alarm (limits monitoring)
Contacts configured as “ALARM” can be ener
-
gized when limits are crossed.
Fail
Contacts configured as “FAIL” will be energized
when a fail situation occurs. Some fail situa
-
tions are automatically signaled by the internal
diagnostics (electronics) of the converter.
Others can be configured by the user (see
section 5-11 Error Configuration). By pressing
the “INFO” button on the main screen the user
is given an explanation as well as a remedy for
the current fail situation. Always connect the fail
contact to an alarm device such as a warning
light, alarm bell or displayed on an annunciator.
“ALARM” Contact “FAIL” Contact
Power Off
NC
NC
Power On
NC
NC
Alarm
NO
NC
Fail
NC
NO
Fail and Alarm
NC*
NO
HOLD
NC
NC
* When a fail situation occurs which is related
to the parameter associated with the contact
(Conductivity, Concentration or temperature)
the contact will go to NC. When the fail
situation is not related to the parameter
associated with the contact the contact will
remain in the state it is currently in.
3-4-3. Contact input
It is necessary to use screening/shielding on
the input signal cables. Terminal 63 is used to
connect the shielding.
3-5. Wiring the mA-output signals
3-5-1. General precautions
The analog output signals of the EXAxt
transmit low power standard industry signals to
peripherals like control systems or strip-chart
recorders (Figure 3-6).
3-5-2. Analog output signals
The output signals consist of active current
signals of 4-20 mA. The maximum load can be
600 ohms on each.
(M4 screw)
Figure 3-8-a. External grounding
Figure 3-8-b. Internal grounding
2
N
1
L
POWER
100-240 VAC/15 VA/ 50/60Hz
FUSE: 500 mA/250 VAC/T
AC
(M4 screw)
Содержание EXAxt 450
Страница 1: ...IM 12D06D05 01E 4th Edition User s Manual Model ISC450G Style S2 Inductive Conductivity Converter ...
Страница 8: ...vi ...
Страница 23: ...15 IM 12D06D05 01E 3 INSTALLATION AND WIRING Figure 3 11 a Figure 3 11 b Figure 3 11 c ...
Страница 54: ...46 IM 12D06D05 01E ...
Страница 55: ...47 IM 12D06D05 01E 10 SPARE PARTS 10 SPARE PARTS See Customer Maintenance Parts List ...
Страница 66: ...58 IM 12D06D05 01E APPENDIX 6 Control drawing for FM approval ...