2. GENERAL SPECIFICATIONS
2.1 Measuring elements
:
Optical dissolved oxygen sensor in probe body and an oxygen exchange cap (OEC).
2.2 Electrical specifications
Output signal:
Bi-directional digital communication
Physical
layer:
RS485
Protocol:
MODBUS, RTU mode
Data rate:
Default 19200 b/s (8, No Parity, 2 stop bits)
Power supply:
+7 … +30 VDC
Max.
1W
in
active
mode
<
150mW
in
stand-by
mode
Cable:
WU10 with one side wired pins and the
other
side
VarioPin
connector
2.3 Mechanical specifications
Case/OEC:
Stainless Steel (1.4435), Ra<0.4μm)
OEC sealings / material:
O-ring EPDM
OEC sealing dimensions mm (inch): ID 9x1 (0.35x0.03)
Shaft sealing mm (inch):
ID 11x3 (0.43x0.11)
Housing size (LxDiam.):
120x12 mm (4.17x0.47)
Weight:
Min. 100g (0.22 lbs)
Sensor mounting system:
Sensor PG13.5 thread
Sensor connection:
VarioPin, 8 pins
2.4 Performance
DO72-W Wide measuring range sensor characteristics
Measuring range:
0 – 22.5 mg/L O
2
Accuracy:
± 0.05 mg/L or 1% of reading
whichever
greater
Repeatability
± 0.05 mg/L or 1% of reading
whichever
greater
Temperature accuracy:
≤
1.0°C (
≤
1.8°F)
1
DO72-T Trace measuring range sensor characteristics
Measuring range:
0 - 2 mg/L O
2
Accuracy:
± 3 ppb or 3% of reading
whichever
greater
Repeatability
± 3 ppb or 3% of reading
whichever
greater
Temperature accuracy:
≤
1.0°C (
≤
1.8°F)
1
Содержание DO72
Страница 1: ...Instruction Manual Model DO72 Optical Dissolved Oxygen Sensor IM12J06J02 00EN P 4th edition...
Страница 14: ...4 DIMENSIONS Fig 8 Dimensions DO72 in mm IM12J06J02 00EN P 14...
Страница 20: ...Fig 16 Example of Final inspection protocol 20 IM12J06J02 00EN P...
Страница 24: ...Subject to change without notice Copyright Printed in The Netherlands 04 2212 IM12J06J02 00EN P...