IM12J06J02-00EN-P
15
5. WIRING
DO72 Sensor cable connection to converter is delivered with wire pin end. For infor-
mation on DO72 sensor cable connection to FLXA402, refer to the user’s manual for
FLXA402 4-Wire Converter (IM12A01F01-02EN).
Fig. 10.
D6 module terminal assignment for DO72 dissolved oxygen sensor
Table 3:
Pin terminal wiring numbering
Fig. 9.
Connection diagram – WU10-V-D (left) and DO72 sensor wiring details (from
FLXA402 manual - right)
Wire Pin no.
VP pin
Wire color
Description
11
G
Yellow
RS485 Data (A)
12
H
Green
RS485 Data (B)
13
D
White (Coax2 shield)
GND
10
C
White (Coax2 core)
7 – 30 VDC
14
S
Black
cable shield / probe body
Remark:
Cables which are not used should be connected to any terminal of some-
where between 15 to 18.