5.1 DO72 optical DO sensor 

grounding

The DO72 dissolved oxygen optical 

sensor housing is made of stainless 

steel with high corrosion resistance 

(SUS 316 L). 

However, electric potential differ

-

ences between the DO72 and a tank 

or other peripherals in the tank can 

cause immediate corrosion through 

electrolysis. Therefore, please 

ensure to avoid electric potential 

differences between DO72 and the 

surrounding environment.

We recommend always to connect 

the DO72 to the protective ground 

at the installation site. Assign the 

device housing and cable shield to 

protective ground.

Especially the length of the cable 

can influence the ground connec

-

tion quality.

If the DO72 is installed in a fitting, 

which is connected to the protective 

ground, then the connection to the 

ground will be established via the 

PG 13.5 stainless steel fitting (see 

Fig. 11). 

If the DO72 is installed in a tank 

which is NOT connected to the 

protective ground (e.g. fitting made 

of plastic or glass), a ground con-

nection should be attached to the 

DO72 PG 13.5 fitting (see Fig. 12). 

You can use a grounding clamp to 

connect the DO72 directly to protec-

tive ground (see Fig. 13 for exam-

ple).

Fig. 11: 

DO72 grounding when installed 

in a tank which is connected to protective 

ground

Fig. 12: 

DO72 grounding when installed in 

a tank which is NOT connected to protective 

ground

Fig. 13: 

Example for a grounding clamp 

that can be used for connecting an DO72 to 

protective ground, when installed in a tank 

without grounding.

PG13.5

Stainless steel thread

VP cable 

shield

Green / 

yellow 

cable

IM12J06J02-00EN-P

16

Содержание DO72

Страница 1: ...Instruction Manual Model DO72 Optical Dissolved Oxygen Sensor IM12J06J02 00EN P 4th edition...

Страница 2: ...hereafter simply referred to as YOKOGAWA when replacing parts or consumables Modification of the product is strictly prohibited The following safety symbols are used on the product as well as in this...

Страница 3: ...se but rather to describe the functional details of the product No part of the Instruction manuals may be transferred or reproduced without prior written consent from YOKOGAWA YOKOGAWA reserves the ri...

Страница 4: ...ental conditions 8 2 7 Regulatory compliance 9 3 INSTALLATION OF DO72 10 3 1 Installation in sensor holder 10 3 1 Preparing the sensor for use 10 3 2 Mounting the sensor 10 4 DIMENSIONS 14 5 WIRING 15...

Страница 5: ...products and parts are guaranteed free from defects in work manship and material under normal use and service for a period of typically 12 months from the date of shipment from the manufacturer Indivi...

Страница 6: ...ction Year code Year Year code Year Year code 2014 P 2026 3 2015 R 2027 4 2016 S 2028 5 2017 T 2029 6 2018 U 2030 7 2019 V 2031 8 2020 W 2032 9 2021 X 2033 A 2022 Y 2034 B 2023 Z 2035 C 2024 1 2036 D...

Страница 7: ...4 m OEC sealings material O ring EPDM OEC sealing dimensions mm inch ID 9x1 0 35x0 03 Shaft sealing mm inch ID 11x3 0 43x0 11 Housing size LxDiam 120x12 mm 4 17x0 47 Weight Min 100g 0 22 lbs Sensor m...

Страница 8: ...g lbs 147 0 32 g lbs 2 6 Environmental conditions Storage temperature 10 to 70 C 14 to 158 F 4 Ingress protection IP X8 according EN 60529 1991 A1 2000 A2 2013 O ring of OEC must be replaced once a y...

Страница 9: ...ndards NEN EN IEC 61010 1 2010 A1 2019 NEN EN IEC 61010 2 030 2010 C1 2011 RoHS2 Directive 2015 863 EU Commission Delegated Directive EU 2015 863 amending Annex II as regards the list of restricted su...

Страница 10: ...tical floating ball holder see Fig 3 For mounting the sensor into the FF20 and FF40 flow fitting use sensor adapter as described in figure 5 6 and 7 3 1 Preparing the sensor for use Make sure that the...

Страница 11: ...4 Adapter for installing DO72 into submersible and floating ball holder Optical dissolved oxygen sensor DOX8HS Submersion type holder Place an O ring here Adaptor for submersible type holder Adaptor...

Страница 12: ...Figure 5 DO72 installation in Yokogawa FF20 fitting6 using the adapter K1520JN K1500DV or K1520JP Adapter K1520JN K1500DV K1520JP IM12J06J02 00EN P 12...

Страница 13: ...n FF40 FS40 fitting6 using the adapter K1520JN K1500DV or K1520JP Figure 7 DO72 Installation in SC4A SS flow fitting6 using the adapter K1523JB K1523JD Adapter K1520JN K1500DV K1520JP Adapter K1523JB...

Страница 14: ...4 DIMENSIONS Fig 8 Dimensions DO72 in mm IM12J06J02 00EN P 14...

Страница 15: ...ssignment for DO72 dissolved oxygen sensor Table 3 Pin terminal wiring numbering Fig 9 Connection diagram WU10 V D left and DO72 sensor wiring details from FLXA402 manual right Wire Pin no VP pin Wire...

Страница 16: ...d in a fitting which is connected to the protective ground then the connection to the ground will be established via the PG 13 5 stainless steel fitting see Fig 11 If the DO72 is installed in a tank w...

Страница 17: ...ves sel Due to a chemical reaction of oxygen with the Na2SO3 the water becomes oxygen free Additional oxygen diffusing from air into the water is removed by surplus Na2SO3 Close the vessel with the s...

Страница 18: ...se a commercially avail able test gas N2 5 0 Lead the gas into a vessel filled with dis tilled water before introducing it in the calibration chamber or a flow through cell with installed DO72 Then fo...

Страница 19: ...utral detergent to the oxygen exchange cap so it can be easi er to remove the dirt Wash the oxygen exchange cap with clean water Before every calibration check that the oxygen exchange cap is not loos...

Страница 20: ...Fig 16 Example of Final inspection protocol 20 IM12J06J02 00EN P...

Страница 21: ...nt on range type Model Suffix code Option code Description DO72 Optical Dissolved Oxygen sensor Range T Trace range 0 2 mg L W Wide range 0 22 5 mg L Insert length 120 120 mm Type AA General purpose C...

Страница 22: ...SLIDE RING 1PC K1523JA Adapters Adapter Pg13 5 in F 40 PPO K1523JC Adapter Pg13 5 in F 40 SS K1520JN Adapter M25x1 5 PG13 5 PVC K1520JP Adapter M25x1 5 PG13 5 RVS K1500DV Adapter M25x1 5 PG13 5 PVDF...

Страница 23: ...gen Exchange Cap membrane compatibility Compatibility Aqueous solutions ethanol methanol No cross sensitivity to pH 1 14 CO2 H2S SO2 ionic species Cross sensitivity to Chlorine gas Organic solvents su...

Страница 24: ...Subject to change without notice Copyright Printed in The Netherlands 04 2212 IM12J06J02 00EN P...

Отзывы: