MODBUS Communication
– Hold Register DO
15
All Rights Reserved. Copyright © 2022, Yokogawa Process Analyzers Europe B.V.
TI 12J06J01-00EN-P
Address
Item name
Data type
Access
Value / description
5517
Calibration Mode
Uint32
R/W
Measurement mode during calibration:
0:
Humid
1:
Dry
D: 0 (for wide range sensors)
D: 1 (for trace range sensors)
5521
Cal0
Float
R/W
Calibration value of Phase shift at the
low oxygen calibration point
L: 30.01
H: 79.99
5523
T0
Float
R/W
Calibration value of Temperature at the
low oxygen calibration point
L: 10.01
H: 29.99
5525
Reserved
Uint32
R
5527
O
2
-2
nd
Float
R/W
Calibration value of Oxygen
concentration (unit defined in HR5535)
at the high oxygen calibration point
L: 0.001
H: 299.999
5529
Cal-2
nd
Float
R/W
Calibration value of Phase shift at the
high oxygen calibration point
L: 10.01
H: 49.99
5531
T2
nd
Float
R/W
Calibration value of Temperature at the
high oxygen calibration point
L: 10.01
H: 29.99
5533
pATM
Float
R/W
Calibration value of Pressure (in hPa) at
the high oxygen calibration point
L: 901
H: 1099
5535
O
2
-2
nd
Unit
Uint32
R/W
0x 0800 0000
% DO saturation
Remark:
1) The unit for the O
2
-2
nd
value, for set
values see HR2089
D: % DO saturation
5703
Measurement Mode
Uint32
R/W
Measurement mode:
0:
Humid
1:
Dry
D: 0
6437
Calibration counter
Uint32
R/W
Counts the number of Low/High Point
Calibrations
D: 0
6457
Serial Number YPA
ASCII
R
Alphanumeric Serial Number code:
16 bytes, 8 addresses