MODBUS Communication
– Hold Register DO
12
All Rights Reserved. Copyright © 2022, Yokogawa Process Analyzers Europe B.V.
TI 12J06J01-00EN-P
Address
Item name
Data type
Access
Value / description
4613
Measurement Temp. max
Float
R/W
The maximum measurement tempera-
ture of the sensor
Remark:
1) Sensor precision is only specified
within the
range 0…+50°C
D: +100°C
4615
Calibration Temp. min
Float
R
The minimum value of the temperature
calibration range
D: 0°C
4617
Calibration Temp. max
Float
R
The maximum value of the temperature
calibration range
D: +50°C
4677
Hours out of Measurement Temperature
Range
Float
R
Accumulated hours in which the sensor
was out of the defined measurement
temperature range (see registers
HR4611 and HR4613)
D: 0
4679
Hours out of Operating Temperature
Range
Float
R
Accumulated hours in which the sensor
was out of the defined operating
temperature range (see registers
HR4607 and HR4609).
D: 0
4895
Pressure Value Raw
Float
R
Measured pressure value in hPa (mbar)
4897
Reference Amplitude Value Raw
Float
R
Reference amplitude raw value in µV
4899
Oxygen Amplitude Value Raw
Float
R
Measured oxygen amplitude raw value
in µV
4901
Oxygen Phase Value Raw
Float
R
Measured oxygen phase shift in degrees
4903
Temperature Value Raw
Float
R
Measured temperature value in the unit
defined in HR2409
4905
Calculated Oxygen Value
Float
R
Calculated Oxygen Value in the unit
defined in register HR2089
4907
Measurement Error
Uint32
R
For details see Section 3.1