MODBUS Communication
– Hold Register DO
11
All Rights Reserved. Copyright © 2022, Yokogawa Process Analyzers Europe B.V.
TI 12J06J01-00EN-P
Address
Item name
Data type
Access
Value / description
4095
Device ID
Uint32
R/W
Modbus address
Remark:
1) Setting is effective immediately
D: 1
L: 1
H: 247
4097
Reserved
Uint32
R
4099
Reserved
Uint32
R
4101
Baud rate
Uint32
R/W
0x 0000 0002
4800 baud
0x 0000 0003
9600 baud
0x 0000 0004
19200 baud
0x 0000 0005
38400 baud
0x 0000 0006
57600 baud
0x 0000 0007
115200 baud
Remark:
1) Setting is effective immediately
D: 19200
L: 4800
H: 115200
4103
Reserved
Uint32
R
4105
Reserved
Uint32
R
4107
Parity
Uint32
R/W
0x 0000 0000
Even parity /1 Stop bit
0x 0000 0001
Odd parity / 1 Stop bit
0x 0000 0002
No parity / 2 Stop bits
Remark:
1) Setting is effective immediately
D: No parity
2 Stop bits
4607
Operating Temp. min
Float
R
The minimum operating temperature of
the sensor
D: -50°C
4609
Operating Temp. max
Float
R
The maximum operating temperature of
the sensor
D: +130°C
4611
Measurement Temp. min
Float
R/W
The minimum measurement tempera-
ture of the sensor
Remark:
1) Sensor precision is only specified
within the range 0…+50°C
D: -40°C