<APPENDIX 1. LIST OF PARAMETERS FOR EACH BLOCK OF digitalYEWFLO>
A1-10
IM 01F06F00-01EN
Relative
Index
Index
Parameter Name
Factory Default
Write
Mode
Explanation
82
2082 NOISE_RATIO
—
—
Indicates the noise balance ratio. When the value of
NOISE_BALANCE_MODE is 1 (Auto), this value cannot
be modified. When it is 2 (Manual), the desired value can
be set as a fixed ratio.
83
2083
SIGNAL_LEVEL
1
O/S
Sets the signal level. Setting range: 0.1 to 20.0
84
2084 FLOW_VELOCITY
—
—
Indicates the current flow velocity in m/s (1061); updated
periodically.
85
2085 SPAN_VELOCITY
—
—
Indicates the span flow velocity in m/s (1061); updated
periodically. (This parameter is set to the value calculated
based on the density under the normal operating
conditions and normal operating temperature specified by
the customer.)
86
2086 VORTEX _
FREQUENCY
—
—
Indicates the current vortex generation frequency in Hz
(1077); updated periodically.
87
2087 SPAN_FREQ
—
—
Indicates the vortex generation frequency at the span
flow in Hz (1077); updated periodically. (This parameter
is set to the value calculated based on the density under
the normal operating conditions and normal operating
temperature specified by the customer.)
88
2088 FLUID_DENSITY
—
—
Indicates the fluid density calculated based on the
temperature data, in the unit determined by DENSITY_
UNIT; updated periodically.
89
2089 SENSOR _ERROR
_RECORD
0
O/S,
AUTO
Record of the following errors regarding the sensor. This
record will be automatically cleared when no error has
occurred for a month.
Recorded errors: Flow over output, Span set error, Pre-
amp fault, EEPROM fault, Sensor fault, Transient noise,
High vibration, Clogging, Fluctuating, Temp over output,
Over temp, Temp sensor fault, Temp convert fault.
Setting range: 0 only. Setting 0 clears the record.
90
2090 MODEL
digitalYEWFLO
O/S,
AUTO
Model of the flowmeter converter
91
2091 ALARM_SUM
0
O/S,
AUTO
Indicates the entire block’s alarm statuses; Disable can
only be set.
153
2153 VOLUME_FLOW
—
—
Indicates the volumetric flow rate.
154
2154 VOLUME _FLOW_
UNIT
—
—
Indicates the unit of VOLUME_FLOW. It links the unit of
XD_SCALE of AI3.
Note 1: The value changes in line with a change to the SIZE_SELECT value or to the unit setting in XD_SCALE of the corresponding AI
block.
Note 2: An intended value which meets both of the following conditions can only be written: CAL_POINT_LO < the intended value, and
SENSOR_RANGE.EU100 > intended value.
Note 3: An intended value which meets both of the following conditions can only be written: CAL_POINT_HI < the intended value, and
SENSOR_RANGE.EU100 > intended value.
Note 4: The flow rate unit can only be written.
Note 5: These parameters are set in accordance with the specifications written in the registered sizing data if it is supplied.
In case of UNCALIBRATION order, they are set to the defaults when shipped.