8-6
IM CA71-E
MF
Queries the measurement function.
When normal
condition
Syntax for query
MF?<delimiter>
⇒
Response: MFm<delimiter>
Description of parameter
m: Measurement function
m=0: 300V AC
m=1: 100V
m=2: 10V
m=3: 1V
m=4: 100mV
m=5: Resistance
m=6: Frequency
m=7: Current
MO
On/Off of measurement/queries the current setting.
Syntax for setting
MOm<delimiter>
Syntax for query
MO?<delimiter>
⇒
Response: MOm<delimiter>
Description of parameter
m: On/Off condition
m=0: Off
m=1: On
MR
Syntax for setting
MRm<delimiter>
Syntax for query
MR? <delimiter>
⇒
Response: MRm<delimiter>
Description of parameter
m: Measuring range
[100V]
m=0: DC
m=1: AC
[10V]
m=0: DC
m=1: AC
[1V]
m=0: DC
m=1: AC
[100mV] (When normal condition)
m=0: 100mV DC
m=1: TcK
m=2: TcE
m=3: TcJ
m=4: TcT
m=5: TcR
m=6: TcB
m=7: TcS
m=8: TcN
m=9: TcL
m=10: TcU
[100mV] (When adjustment)
m=0: 100mV DC
m=1: TcK
m=2: TcE
m=4: TcT
[Ω]
m=0: 400Ω
m=1: Pt100
m=2: JPt
[Freq]
m=0: 100Hz
m=1: 1kHz
m=2: 10kHz
[mA]
m=0: 20mA
m=1: 100mA
When normal
condition/adjustment
When normal
condition/adjustment
Sets the measuring range/queries the current setting.
8.7 Detailed Description of Commands