IM CA71-E
10-7
Method of Calibrator
Adjustment
1
2
3
4
5
6
7
8
9
10
11
12
13
App
10.3 Adjusting Measurement Functions
Table 10.2 Adjustment Setpoints of Measurement Functions
Range
Adjustment Points *
2
Remarks
CAL 0
CAL FS
DC 100 mV
---
100 mV
---
DC 1 V
---
1 V
---
DC 10 V
---
10 V
---
DC 100 V
---
100 V
---
DC 20 mA
---
20 mA
---
DC 100 mA
---
100 mA
---
400 Ω
0 Ω
380 Ω
Three-wire connection
AC 1 V
0 V
1V/50 to 60 Hz
---
AC 10 V
0 V
10 V/50 to 60 Hz
---
AC 100 V
0 V
100 V/50 to 60 Hz
---
AC 300 V
0 V
300 V/50 to 60 Hz
---
*2: Apply the reference input signals listed above from the calibration standard.
TIP
• You can also select only the range in need of readjustment to adjust
it separately.
• Always make zero-point (0) adjustments together with full-scale (FS)
adjustments.
10.3.1 Adjusting DC Voltage and DC Current Ranges
Step 1:
Press the
key while simultaneously holding down
the
key. The LCD shows
.
Step 2:
Pressing the highest-order key causes the LCD to show
.
Step 3:
Pressing the
key enters the measurement CAL mode.
The
indicator blinks on the LCD and the symbol
appears.
The calibrator is now ready for the defining the CAL FS
setpoint of measurement functions.
10.3 Adjusting Measurement Functions