Step 5 -- Master Event Position
This section describes the operation to set the event position (master) to be used as the
reference for making a report.
The master event is set up by either of [File Open] and [Auto Generation].
If a measurement is made bidirectionally, the master events must be set up for both
directions respectively.
The master setup by [Auto Generation] is used as an example in the following explanation.
c
Click [Auto Generation].
The [Event Auto Generation Setup] window will appear.
d
Select [Generation from Auto Search].
Set up Search Condition and/or Analysis Condition as
necessary.
e
Click [OK].
See the help facilities.
→
Trace Output
→
Step 5
→
Auto Generation
f
If a measurement is made bidirectionally, the master events
must be set up for both directions respectively.
Click [Reverse] in the [File] area, and then perform Steps
c
through
e
.
4
5
7
g
Click [Total Loss Marker Edit].
(Not available if this setup is not necessary due to
the total loss setup in the [
Step 3 -- Layout Options
(Trace Output Structure)].)
h
Place the marker
N
at the start
point of the fiber, and the
marker
F
at the endpoint.
i
Click [Output
Trace Scale
Edit].
j
Set the desired vertical/
horizontal scales and the
waveform display position.
k
Click [Next].
19