34
IM AQ7280-02EN
Removing the Optical Power Meter or Visible Light Source Module
1.
Remove the rubber cover from the optical power meter/visible light source module slot of the OTDR mainframe.
2.
Loosen the optical power meter or visible light source module attachment screw with a coin or a flat-bladed screwdriver until the
screw head moves up and down. The screw is designed so that it does not come off the module.
3.
Hold the screw and pull the optical power meter or visible light source module up from the OTDR mainframe.
4.
Attach the included cover bracket. Tighten the screw (M4) with a Phillips screwdriver.
The proper tightening torque is about 1.2 N-m.
Cover bracket
Screw
Note
If you are not going to install an optical power meter or visible light source module, attach the cover bracket to prevent foreign
substances from entering the slot.