34
IM AQ6375-02EN
9
Connecting the DUT
Procedure
Connecting Optical Fibers
1.
Clean the tip of the optical fiber with a fiber cleaner.
2.
Open the instrument’s optical input connector cover.
3.
Connect the optical fiber’s optical connector to the optical input connector on the
instrument.
CAUTION
• Before connecting an optical fiber to the instrument, make sure that the start-up
initialization process has finished. If a strong light source is input during start-up,
the optical section can be damaged.
• Be sure to clean the tip of the optical fiber’s optical connector before connecting.
• Do not try to forcefully attach the optical fiber’s optical connector with the plug
inserted at a slanted angle. Doing so may damage the instrument’s optical
connector’s components or the connector itself.
• Before connecting the input light, make sure that it does not exceed the
AQ6375’s maximum rated level. If input light exceeding the maximum rated level
is introduced, the optical section may be damaged.
• Press the optical connector hard against the cleaning surface of the special
cleaner to clean it. If it is not pressed hard against the cleaning surface, it may
not be possible to properly clean the optical connector.
Connecting the DUT (Light Source)
4.
Clean the top of the optical connector on the other end of the optical fiber with a
fiber cleaner.
5.
Connect the optical connector on the other end of the optical fiber to the optical
connector on the DUT.
Measuring System
AQ6375
Light source
Optical fiber