Connecting the Power Cord
1.
Check that the MAIN POWER switch on the rear panel is off.
2.
Connect the power cord plug to the power inlet on the rear panel.
3.
Connect the other end of the cord to an outlet that meets the following conditions. Use a
grounded three-prong outlet.
Item
Rated supply voltage*
100 VAC to 240 VAC
Permitted supply voltage range
90 VAC to 264 VAC
Rated power supply frequency
50/60 Hz
Permitted supply frequency range
48 Hz to 63 Hz
Maximum power consumption
Approx. 100 VA
* This instrument can use a 100 V or a 200 V power supply. The maximum rated voltage
differs according to the type of power cord. Check that the voltage supplied to the
instrument is less than or equal to the maximum rated voltage of the power cord that you
will be using before use.
Three-prong outlet
Power cord
OFF
MAIN
POW
ER
ON
100-
240V
AC
150V
A
MAX
50/60
Hz
FUSE
250V
T 5A
CAUTION
• When turning the power on, do not apply a high-output light source to the instrument. If you
do, the optical section may be damaged.
• Do not operate the POWER or MAIN POWER switch during the initialization process.
Doing so may damage the instrument.
French
ATTENTION
• Lors de la mise sous tension, n’appliquez pas une source de lumière à haute performance
à l’instrument. Si vous le faites, la section optique pourrait s’endommager.
• N’appuyez pas sur les interrupteurs POWER ou MAIN POWER pendant le processus
d’initialisation. Cela pourrait endommager l’instrument.
Turning the Main Power Switch On
1.
Flip the
MAIN POWER
switch on the rear panel to the ON ( | ) position.
The POWER switch on the front panel illuminates in orange.
OFF
MAIN
POW
ER
ON
100-
240V
AC
150V
A M
AX
50/6
0Hz
FUSE
250
V T
5A
2.3 Connecting to the Power Supply and Turning the Power Switch On and Off