Operation

¡

¡

¡

¡

¡

Normal Measurement

[Starting Normal Measurement]

q

Press  MEASURE  button to turn on the power supply.  A laser
beam is emitted and the measurement starts.  (In the Normal Mea-
surement mode, the unit takes measurements while  MEASURE
button is pressed.)

w

Point the laser beam at a measuring object and aim it at the
center of the area to be measured.  For the distance from this
thermometer to the measuring object, refer to page 9.

e

In the Normal Measurement mode, the measurement indication

 " is shown on the LCD display and a laser indicator blinks.

[Quitting Normal Measurement / HOLD mode]

q

Release  MEASURE  button.  Then, the laser beam turns off and
a value measured last is displayed for 15 seconds (HOLD mode).
The power supply turns off automatically after 15 seconds.

* NOR HOLD can be changed into MAX/MIN HOLD.
(See "Setting" on page 15)

¡

¡

¡

¡

¡

Continuous Measurement

[Starting Continuous Measurement]

q

Press and hold both of  MEASURE  button and  SET/ENTER  button simultaneously for two seconds or
more, then the Continuous Measurement mode switched on.  (In the Continuous Measurement mode, the
thermometer keeps measuring temperatures without pressing  MEASURE  button.)

w

During the Continuous Measurement mode, LCD display shows
"

".

[Quitting Continuous Measurement]

q

To quit the Continuous Measurement mode, press and hold  SET/ENTER  button for two seconds or
more.  Then, it switches to HOLD mode.  The power supply turns off automatically after 15 seconds
in the HOLD mode.

CAUTION : The laser beam is constantly emitted during Continuous Measurement mode.

Be careful of it while operating the thermometer.

Set the battery supplied as an accessory and operate the unit according to the following procedure:

8

53008-E  12.5.12 1:12 PM  ページ 7

Содержание 53008

Страница 1: ...product in a place except for home Registration No Equipment Name Trade Name Manufacturer Country of Origin KCC REM IMY EEN326 Emission Thermometer Yokogawa Meter Instruments Corporation Yokogawa Met...

Страница 2: ...surface of an ob ject into temperature This thermometer measures the surface temperature of solid and liquid without contacting them The temperature of gas cannot be measured by this thermometer Cont...

Страница 3: ...sult in hazardous radiation exposure Safe Usage When operating the instrument be sure to observe the cautionary notes given below to ensure correct and safe use of the instrument If you use the instru...

Страница 4: ...alue is near 0 reflect surrounding tempera tures As this thermometer s sensitiv ity to emissivity is fixed at 0 95 0 70 the displayed temperature could differ from the actual temperature of objects wh...

Страница 5: ...0 Accuracy 0 95 0 to 200 C 2 C 201 to 500 C 1 Repeatability 1 C of reading value Display resolution 1 C Sighting method Non coaxial laser marker Class 2 HOLD time 15 seconds Continuous measurement mod...

Страница 6: ...e is no button to turn ON OFF the power supply Pressing MEASURE button wakes up the thermometer The power supply turns off automatically after 15 seconds absence of pressing any button Name of Compone...

Страница 7: ...s a Class 2 laser prod uct as defined by IEC60825 1 Safety of Laser Products Part1 Equipment classi fication and requirements In addition this instrument complies with 21 CFR 1040 10 except for deviat...

Страница 8: ...matically after 15 seconds NOR HOLD can be changed into MAX MIN HOLD See Setting on page 15 Continuous Measurement Starting Continuous Measurement qPress and hold both of MEASURE button and SET ENTER...

Страница 9: ...away from the measuring object unless there is any obstacle However please note that the measuring field of view enlarges in proportion to the measuring distance This thermometer has an optical resolu...

Страница 10: ...ess and hold CALL button for two seconds or more in HOLD Mode refer to page 8 and CALL sign appears on the LCD display In CALL recalling mode the last record of tem perature data and the corresponding...

Страница 11: ...page 14 Record Erasing HOLD mode Selections refer to page 15 rWhen each setting is completed it automatically turns to the next setting mode For example when setting of High Limit Temperature for Alar...

Страница 12: ...Limit Temperature for Alarm cannot be set at lower temperature than Low Limit Temperature for Alarm Turning ON OFF Alarm Function qWhen the High Limit Temperature setting is completed the setting mod...

Страница 13: ...the displayed value as the Low Limit Temperature for Alarm Note The Low Limit Temperature for Alarm cannot be set at higher temperature than High LimitTemperature for Alarm Turning ON OFF Alarm Funct...

Страница 14: ...f data Emissivity Emissivity is a value that indicates the infrared energy emitted from the surface of an object Every object has its own emissivity value and it varies depending on the surface condit...

Страница 15: ...The measured object is smaller than the field of view Adjust the measuring distance referring to the Field of view described in page 9 The thermometer is affected by a nearby heating source Cut off t...

Страница 16: ...rmation or damage Periodical Inspection It is recommended that the thermometer be calibrated annually For further information please contact the vendor from which you purchased the product Battery Bat...

Страница 17: ...17...

Страница 18: ...18...

Отзывы: