— 7 —
IM 3201–E
5. When it is desired to measure a voltage of high frequency,
connect the
COM-
terminal to the ground side of the measuring
circuit. This arrangement prevents indication errors caused by
the stray capacitance between the tester and the ground.
(Refer to
Fig 3.1 Frequency Characteristics
in which an
example of the frequency characteristics is shown.)
6. When measuring AC voltage in the circuit where a DC component
is included, connect the leads respectively to the
OUTPUT
terminal
and
COM-
terminal. For further procedures, follow instructions
described in the paragraph “AC Voltage Measurement”.
7. To the
OUTPUT
terminal a 0.2 μF capacitor (400 W. V) is
internally provided in series so that the DC component is
impeded and only the AC voltage is indicated.
8. In case of low frequency, voltage drop across the capacitor
causes an error in indication. This error depends upon the
measuring range. (Refer to
Fig. 3.2 Frequency Characteristics
when OUTPUT Terminal is Used
)
0.2
0
1
2
3
4
5
0.5
1
2
10
5
50
100
20
Frequency (kHz)
Error
(%)
120 V Range
30 V Range
3 V
Ra
nge
12 V
Ra
nge
Fig 3.1 Frequency Characteristics
50
100
200
500
1000
20
Frequency (Hz)
-10
0
-20
-30
-40
Error
(%)
3 V Range
12 V
Range
30 V Range
Fig. 3.2 Frequency Characteristics
when OUTPUT Terminal is Used