6
Sensors:
Temperature
Range:
-2 - 50ºC
Accuracy:
±0.05ºC
Resolution:
0.01ºC
Type:
pt 100 platinum element
Conductivity
High Range:
0 - 80 ms/cm
Accuracy:
±0.05 ms/cm
Resolution:
0.02 ms/cm
Low Range:
0-8000 us/cm
Accuracy:
±5 us/cm
Resolution:
3 us/cm
Type:
Four electrode cell
Salinity
Range:
0 - 60 ppt
Accuracy:
±0.05 ppt
Resolution:
0.02 ppt
Type:
See Appendix 1 - Conversions Used
Dissolved Oxygen
Range:
0 - 200% saturation
0 - 20 mg/l
Accuracy:
±0.5%
Resolution:
0.1%
Type:
Active silver and lead electrode sensor with PTFE* membrane and built-in stirrer
Turbidity
Range:
0 - 600 ntu
Accuracy:
±0.5 ntu (0 - 300 ntu range)
±5 ntu (300 - 600 ntu range)
Resolution:
0.3 ntu
Type:
Nephelometric measurement from a 90
°
sensor with pulsed infra-red light source
Содержание 615