Appendix 1 - Conversions Used
The YK615 measures dissolved oxygen as % saturation and then automatically converts
the reading to milligrams per litre. This conversion is calculated from the dissolved
oxygen solubility tables found in International Oceanographic Tables vol.2. National
Institute of Oceanography 1972.
The conversion between conductivity and salinity is performed using the Practical
Salinity Scale. UNESCO Technical Papers in Marine Science 1983.
The conversion of low conductivity raw data to conductivity referenced to 25 Deg C is
performed using constants derived from HANDBOOK OF CHEMISTRY AND PHYSICS,
1963, Chemical Rubber Publishing Company, Page 2691, Conductivity of Standard
Solutions using KCl, 0.001 M solution.
Redox potential (ORP) conforms to International Standard IEC 746-5, “Expressions of
Performance of Electrochemical Analyzers, Part 5: Oxidation-Reduction potential”. In
accordance with this standard, the Redox potential is referred to the standard
(“normal”) hydrogen electrode (NHE) and is expressed in mV.