Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
This CCQ-800 is intended for use in the electromagnetic environment specified below. The customer or the user
of the CCQ-800 should assure that the device is used in such an environment.
Immunity test
IEC60601 Test level Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
Portable and mobile RF communications
equipment should be used no closer to any part
of the CCQ-800, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m