19
2) Perform alignment
Move the lens horizontally to align the target with vertical line of the cross, and then move the
lens vertically to align the target with horizontal line of the cross until the target becomes blue,
the procedure is as follows:
●
Arrow indicates the direction should be moved toward, move the lens slowly during alignment
in the direction of arrow.
●
The lens should be always contacted with support, when move the lens, keep the frame adjoin
to push board.
3) Move the lens in vertical and horizontal direction slightly until the measured date stabilizes,
then the message "Near Measurement Complete" appears.
f. Measure the other lens
g. Print out measured data
●
After the ADD power is measured, the automatic L/R selection function begins to work.
●
The measurement results are for reference only.
6.6 Measuring Contact Lens
Detailed steps of contact lens measurement are shown as below under Contact Lens
Measurement Mode:
a. Change the spectacle lens support into contact lens support especially for contact lens
b. Contact Lens Measurement Mode switch
Come to Parameter Setting interface, and set parameter of “Contact Lens” to "ON" or "ONLY",
then save the result to finish Contact Lens Measurement Mode switch process (see details in
6.16).
c. Set contact lens
Set the lens onto Lens support with convex side up. If it's soft contact lens, remove moisture
from the surface with soft cloth before putting it onto Lens support.