background image

        

USER MANUAL 

TYPE ML-2012/13

    

Manufacturers of low power instruments 

 

page 26

 

We  recommend  the  use  of  the  USB-connection  over  the  wireless  function.  This  is  because  of  possible 
drop outs in communication. When a firmware upgrade procedure is interrupted, the upgrade will fail, but 
the unit continues to operate, with the previous version. The USB- connection is faster and more reliable. 

4.10.3  Firmware upgrade over the air 

 
It is also possible to perform a firmware upgrade over the air. For this you must use the software-package 
“terminal link”. It allows you to modify or upgrade your data logger from remote. For more information 
consult your local ydoc dealer. 

4.10.4  Firmware Driver limitations 

 
The  ML-2012/13  Data  logger  is  equipped  with  various  drivers,  for  several  tasks.  The  number  of  total 
drivers  is  limited  to  16.  This means  you  can  choose  maximum  16  drivers  from all  the  drivers  available. 
Each  driver  can  handle  (collect  or  send)  a  number  of  parameters.  The  total  number  of  parameters  is 
limited to 64. So, theoretically, the ML-2012/13 can handle up to 16 sensors with each 4 parameters. But 
in  this  case  there  is  no  output  possible.  In  most  cases  two  or  three  drivers  are  needed  for  minimum 
operation  (internal  driver,  tcp  output-driver,  or  email-driver).  When  a  user  has  an  additional  analog 
sensor, it will take a driver as well.  
 

Example: 

A user wants to connect a large number of INW CT2X sensors. This sensor measures temperature and 
conductivity.  So  each  sensor  “uses”  two  parameters.  Also  he  likes  to  send  the  data  via  FTP,  and  he 
enables  the  following  internal  sensors:  Rest  Capacity,  Battery  Voltage,  and  Current.  So,  besides  the 
Sensor drivers there are 2 drivers needed: 
 

Internal Driver

 

Output Driver

 

 

So the maximum number of Sensor-drivers is: 16 – 2 = 14 
So there are 14 Sensor-drivers left for use with the CT2x sensors. 
The total number of parameters is: 14 * 2 + 3= 31  
So the number of parameters is no problem (31 < 64) 
He can connect 14 CT2X sensors max. 

4.10.5  Power Switch Limitations 

 
The Power Switch is capable of powering sensors up to 200 mA. The output voltage is 12 Volts DC. 

4.10.6  Modem 

 
The  modem  inside  the  ML-2012/13  (the  smallest  modem  in  the  world,  @  the  moment  this  manual  is 
written),  is  capable  of  GPRS  and  GSM  communications.  It  has  quad  band  support,  so  it  can  be  used 
worldwide. For selecting the proper frequency band, use the “modem settings” in the configuration menu. 
In  most  cases  GPRS  communication  is  requested,  so  use  a  GPRS  enabled  SIM-card  to  inhibit  this 
operation.  Consult  your  local  telecom  provider  for  the  right  SIM-Card  and  –  settings.  If  GSM 
communication is requested, use a GSM-DATA subscription. GSM communication is suitable for modem 
dial in communication. We strongly advise to activate this option in the SIM cards used in the logger. This 
is often used to re-configure a logger, remotely, or to monitor actual values. In most cases, normal GPRS 
communication is sufficient for the operation of the data logger. 
 

4.10.7  Modem Firmware Upgrade 

 

Содержание ML-2012

Страница 1: ...USER MANUAL TYPE ML 2012 13 Manufacturers of low power instruments page 1 Type ML 2012 13 Low power GPRS data logger Title User Manual ML 2012 13 Date 30 Oct 2014 Version 1 4...

Страница 2: ...12 13 data logger prepares you to install it and tells you how to put it into operation The Chapter Operating Basics covers basic principles of operation of the data logger The operating interface men...

Страница 3: ...4 6 Analog Differential Inputs ML 2013 only 20 4 6 1 Differential input ports theory of operation 20 4 6 2 Common mode noise rejection 20 4 6 3 Using Load Cells With the ML 2013 20 4 6 4 Bridge of Whe...

Страница 4: ...tion Battery Life 33 5 Pin configuration 34 5 1 ML 2012 34 5 2 ML 2013 35 5 3 Pin Description 36 5 3 1 Analog Inputs 36 Analog Input 1 to 4 36 Analog Input 5 to 8 ML 2013 only 36 5 3 2 RS 485 A B 36 5...

Страница 5: ...be used to connect pyranometers or in combination with a stable excitation voltage to connect Wheatstone resistive bridge sensors like load cells The data logger is provided with generic serial port d...

Страница 6: ...the logger to an external VDC source ML 2013 AC As ML 2013 but with 85 265V AC adapter integrated in cover To connect the logger to an AC power grid ML 2012 13 PCB As ML 2012 13 but without enclosure...

Страница 7: ...de PA66 GF20 with in between a silicon gasket The ML 2013 can be used outdoors 1 Analog inputs 2 RS232 connector 3 Excitation DC DC convertor 4 Digital inputs 5 SDI 12 power switch and Excitation conn...

Страница 8: ...he menu the configuration and network settings must also mach to those supplied by the SIM card provider The pin code of the SIM card must be removed prior to insertion in the data logger To prevent p...

Страница 9: ...on both devices ESD discharges could cause invisible damage This endangers long term stability and proper operation 2 3 Power on for the First time In the factory the data logger is programmed with th...

Страница 10: ...of the firmware NOT the wiring For wiring information see the reference Example Let s Configure a ML 2012 13 for operation with the following Analog pressure transmitter 4 20mA CT2X conductivity tempe...

Страница 11: ...Next type 7 for Analog sensors Choose 1 for Port 1 mA and the next screen will appear General settings 0 Exit 1 System Name YDOC 2 Data logging interval Normal 00 01 00 Alarm Not Used 3 Direct Data ou...

Страница 12: ...urement at 1m water level and a measurement at 2m water level while the sensors range is 0 10m 6 If necessary option 9 can be used to perform an offset correction by measuring a single calibration poi...

Страница 13: ...e interval of digital sensors too short 1 second is possible but probably not sec Just to be sure we choose 10 sec Consult the manual of your digital sensor and check the response time The sample inte...

Страница 14: ...ddress So make sure this address is valid Now perform an Email test and check if it is working right It is strongly recommended to include some internal sensors in the configuration because of the mon...

Страница 15: ...V Ctrl D The result should be like this Internal sensors 0 Exit 1 Name Internal 2 Sample interval Normal 00 00 01 Alarm Not Used 3 Battery Capacity mAh 17000 4 Battery replaced Yes 5 Rest Capacity Res...

Страница 16: ...is in the active state When the data logger is in sleep mode the tasks triggered by the sample interval will NOT execute 4 1 2 Data logging interval This interval determines when a data value obtaine...

Страница 17: ...user can evaluate these values every 5 seconds 4 2 SDI 12 The ML 2012 13 is provided with an SDI 12 port This port is connected to port 2 and is shared with the normal RS232 operation This means that...

Страница 18: ...ast character it switches to RX mode immediately without the delay Therefore no replies are missing Your ML 2012 13 has a true SDI 12 port For more information on the SDI 12 protocol see www sdi 12 or...

Страница 19: ...as a master and retrieves the information from the slaves Only one slave can respond to the requests of the master at a time To set up your RS485 sensor for use with the ML 2012 13 make sure that the...

Страница 20: ...s are very sensitive and particularly suitable for measuring signals from load cells 4 6 1 Differential input ports theory of operation Differential inputs are very convenient for measuring differenti...

Страница 21: ...table reverence Voltage often called Excitation Voltage The ML2013 has an Excitation output for this purpose 4 6 5 Excitation Output The ML 2013 has an Excitation output for powering load cells or oth...

Страница 22: ...is adjusted For input 5 this is very accurate it will be 0 05 For an application that requires an even higher accuracy the user can and should do a user overall calibration 4 6 9 Overall Calibration...

Страница 23: ...r systems or NPN outputs 4 7 2 Pull down type The pull up type of input means that there is an internal resistor mounted between the input and the ground So when NO signal is connected the input will...

Страница 24: ...his alarm level is reached when the data logger encounters a value which is higher than the high high Limit this is the 2nd and most urgent state of alarming This type of alarming is used for very rar...

Страница 25: ...ort 4 10 1 When to use Firmware upgrades Normally a user never uses this feature as long as he is satisfied with the performance of the system In time however the need for additional features may aris...

Страница 26: ...f INW CT2X sensors This sensor measures temperature and conductivity So each sensor uses two parameters Also he likes to send the data via FTP and he enables the following internal sensors Rest Capaci...

Страница 27: ...card adaptor or can be downloaded from the card by using the menu option Data download We recommend not removing the SD card from the logger 4 11 1 Inserting an SD card At the factory the SD card is a...

Страница 28: ...So the D record consists of a D character followed by a timestamp and after that one or more series of parameter code and parameter name So the D record consists of these elements 1 D 2 Parameter Cod...

Страница 29: ...m Message A short and comprehensive message to explain the type of problem encountered Additional System Message Some more information on the exception Example S records 4 12 3 Data Modifiers Normally...

Страница 30: ...ming a lot of power so try to minimize this Consult the manual of the sensor for warm up times A warm up time of 5 minutes is possible and can be used in rare situations but the battery life will be s...

Страница 31: ...other measurement For safety reasons a user can include some internal measurements in the configuration to allow monitoring of the performance of the system Example configuration Rain Measurement Here...

Страница 32: ...Quantity Quantity is the difference between the actual counter value and the previous counter value So when your data log interval is set to 10 minutes this parameter shows you the amount of pulses p...

Страница 33: ...the data logger this power monitor keeps track of the power consumption of the device When the device is going into sleep mode a fixed value is used to calculate the power consumption Both are calcula...

Страница 34: ...ve terminal analog input 3 4 0 20mA input 4 Positive terminal analog input 4 5 8 GND Ground X4 1 SDI 12 0 5V Terminal to connect SDI 12 sensors 2 Switched power output 12V 100mA Terminal to supply pow...

Страница 35: ...input 6 0 80mV Positive terminal analog input 6 8 diff input 6 0 80mV Negative terminal analog input 6 X4 1 Switched module power output 5V Terminal to switch on external modules a o display 2 SDI 12...

Страница 36: ...5 3 3 Power Switch This is an output to drive one or more sensors It holds a level of 12 Volts and is capable of driving up to 200 mA 5 3 4 VBAT This is the main power supply input for the board The...

Страница 37: ...re the antenna is in vertical position as the GSM and GPRS radio signals are vertically polarize the antenna should be vertical positioned for maximum performance Do not place the antenna near metal s...

Страница 38: ...fe stability see spec sheet Also when operating in a harsh environment the need for a recalibration can be applicable YDOC can perform overall calibrations any time you like 6 3 XRAY In the uncommon e...

Страница 39: ...disposed Regulations for disposal of batteries may be different Remove the batteries and dispose them in accordance to the local regulations for batteries 9 Transport and Storage The following requir...

Страница 40: ...solution 12 bits Input type 0 20 mA Channel 1 to 4 Number of Differential Channels 2 ML 2013 only Resolution 16 bits Input type Voltage 0 2000 mV 8 different ranges Smallest detectable step 1 uV Memor...

Страница 41: ...USER MANUAL TYPE ML 2012 13 Manufacturers of low power instruments page 41 11 EU Declaration of Conformity 11 1 ML 2012...

Страница 42: ...USER MANUAL TYPE ML 2012 13 Manufacturers of low power instruments page 42 11 2 ML 2013...

Страница 43: ...RS232 485 SDI 12 Generic SDI 12 NMEA GGA Generic RS232 NMEA DBT Generic RS232 NMEA HDG Generic RS232 NMEA MWV Generic RS232 AWAC Current Nortek RS232 AWAC Wave Nortek RS232 Thalimedes OTT RS232 SDI12...

Страница 44: ...y wrong it must be adjusted If it is totally out of date it designates a RTC problem There should be NO errors on this startup If there are errors contact your local YDOC dealer A screen dump of the s...

Страница 45: ...end us an email with your comments Thank you for your assistance Manual Name ML 2012 13__________________________________________________________ Revision Number 1 0___________________________________...

Отзывы: