Normal Operations
Resistivity display will flash during the 2-hour initial rinse. After completion, the system enters
Polishing mode for 10 minutes in which the inlet solenoid is open, and the pump is on. The UV
bulb is also on if equipped. UV LED light by the d
ual conditioning cartridge will provide feedback
when the UV bulb is on. Polishing mode recirculates water internally to maintain water purity. The
system continuously monitors water quality, system performance and consumable health. System
parameters for recirculation can be adjusted. See “Recirculation Adjustment”.
Flashing Final Resistivity during 2-hour “Initial Rinse”
Final Resistivity during factory recirculation setting
NOTE:
Initial rinse can be bypassed to enter Polishing mode by holding the Initial Rinse button
at the bottom of the display for 3 seconds. This is useful if the system is restarted with previously
rinsed consumable packs.
21
OPERATION PROCEDURES
Содержание WA Series
Страница 5: ...3 MODEL TYPES...
Страница 49: ...47 DIMENSIONAL DRAWINGS...
Страница 50: ...48 DIMENSIONAL DRAWINGS...
Страница 51: ...49 PRE INSTALLATION CHECKLIST...