15
Operation Method
Apparatus/Liquid Sterilization Course
After the sterilization operation, turn the lower
exhaust valve knob to the ON position (Figure
24) to discharge the steam in the cylinder for
pressure relief. When the temperature in the
cylinder is lower than about 102
℃
, the interlock
lamp turns to blink (Figure 25), indicating that
the pressure in the cylinder is basically
eliminated.
After the pointer of the pressure gauge returns
to zero (Figure 26), turn the cover CCW and
separate the upper and lower handles
completely (Figure 27). At this point, the
interlock lamp goes out (Figure 28), indicating
that the cover can be removed and the
sterilization articles can be taken out. (This
figure is for reference only, the pressure gauge
is subject to the real object)
Note
: The outside surface of this unit may get hot to the touch contact between unprotected skin and
outside surface of this unit during operation may cause burns. Always use protective gear to shield
operator from hot surfaces
Figure 24
Figure 25
Figure 26
Figure 27
Figure 28
Exhaust
Drain