Yamato SI 401 Скачать руководство пользователя страница 22

 

 

19 

 

Operation Method 

Quick Auto Stop Operation 

 

Quick auto stop 
  operation procedure 

 

 

 

This operation is used to specify the period up to automatic stop, 
i.e., sets the auto stop timer during operation. 

1. Set the time up to stop during fixed temperature operation 

  Check that the FIXED TEMP lamp lights on and that the unit is 

under operation. 

  Press the TIMER key. 

  The measurement temperature display screen displays the 

character "tim", which indicates the timer setting.    The setting 
temperature display screen displays the current setting time with 
blinking. 

  Select the time by pressing the "

▼▲

". 

Timer function: 

  The maximum setting time is "999 hours and 50 minutes". 

  The  time  can  be  set  in  increments  of  a  minute  under  99  hours 

and 59 minutes. 

  It can be set in increment of ten minutes over 100 hours. 

  The "

▼▲

"can change the setting time quickly when it is pressed 

continuously.    Press them discontinuously when fine adjustment 
is needed. 

 

 

2. Start timer operation 

  Press  the  START/STOP  key  for  one  second  after  deciding  the 

time. 

  Timer  operation  starts  with  the  FIXED  TEMP  and  AUTO  STOP 

lamps lighting on. 

  The timer is activated at the point when the START/STOP key is 

pressed. 

 

 

3. Stop/terminate timer operation 

  The operation stops automatically at setting time. 

  Buzzer  continues  to  sound  for  about  five  minutes  at  operation 

stop. 

  The  setting  temperature  screen  displays  the  character  "End", 

which  indicates  termination  of  operation,  with  the  FIXED  TEMP 
and  AUTO  STOP  lamps  lighting  on.    Press  the  START/STOP 
key  to  terminate  the  timer  operation  mode.    The  screen  returns 
to the initial setting screen. 

To correct or check setting… 

Changing  the  setting  temperature  during  operation  is  possible  by  pressing  the  FIXED  TEMP  key.    Press 
the ENTER key after changing the setting. 

Changing the setting temperature during operation is available by pressing the FIXED TEMP key.    Press 
the ENTER key after changing the setting. 

Press  the 

  key  to  display  the  setting  temperature,  operation  mode  and  residual  time  on  the  setting 

temperature screen. 

Содержание SI 401

Страница 1: ...nk you for purchasing Drying Sterilizer SH Series of Yamato Scientific Co Ltd To use this unit properly read this Instruction Manual thoroughly before using this unit Keep this instruction manual arou...

Страница 2: ......

Страница 3: ...Setting of Overheating Prevention Device 15 Sterilizing Operation 16 Fixed Temperature Operation 18 Quick Auto Stop Operation 19 Auto Stop Operation 20 Auto Start Operation 22 Program Operation 24 Oth...

Страница 4: ...a problem that may cause a serious accident or ev en f atality CAUTION If the caution is ignored there is the danger of a problem that may cause injury damage to property or the unit itself Meaning o...

Страница 5: ...ution electrical shock Caution scald Caution no road heating Caution not to drench Caution water only Caution deadly poison Prohibit Prohibit generally Prohibit inflammable Prohibit to disassemble Pro...

Страница 6: ...it is dangerous and not recommended Do not use the power cord if it is bundled or tangled Do not use the power cord if it is bundled or tangled If it is used in this manner it can overheat and fire m...

Страница 7: ...lectric shock may be caused Do not use a branching receptacle which may cause the heat generation 2 Choose a proper place for installation Do not install this unit in a place where Rough or dirty surf...

Страница 8: ...Never use this unit in an area where there is flammable or explosive gas This unit is not explosion proof An arc may be generated when the power switch is turned ON or OFF and fire explosion may resul...

Страница 9: ...xpectible trouble or malfunction CAUTION 6 Do not make an overload 7 Do not set samples in close formation The withstand load of shelf is 15kg uniform load Set the samples apart each other The tempera...

Страница 10: ...on To prevent take measures that putting up a notice of operating etc 10 Handling of power code Two shelves are attached on the product One of them is previously fixed on the lowermost stand of bracke...

Страница 11: ...he cord s insulation may melt and cause a fire or electrical shock If the power cord becomes damaged wiring exposed breakage etc immediately turn off the power at the rear of this unit and shut off th...

Страница 12: ...9 Description and Function of Each Part Main Unit Temperature sensor Main switch Earth leakage breaker Control panel Door Door handle Power code Exhausted opening SI401 601 Front view Rear view...

Страница 13: ...TO STOP Lamp Blinks while setting quick auto stop timer or auto stop timer Lights while quick auto stop timer or auto stop timer is running AUTO START Lamp Blinks while setting auto start timer Lights...

Страница 14: ...gram Operation Sv 1 Program Temperature Setting Used for setting the temperature for each step in the program Sv 1 to Sv 30 is shown t 1 Program Time Setting Used for setting the time for each step in...

Страница 15: ...ter specified period 19 4 Auto Stop Operation This operation is used to specify the automatic stop time in the fixed temperature operation Pressing the TIMER key displays AS p The setting temperature...

Страница 16: ...again when the POWER switch is pressed again manual reset 2 Calibration offset function This calibration offset function is for calibrating the difference occurred between the required in furnace temp...

Страница 17: ...14 Operation Method Operation Mode Function Setting Key and Characters The operation mode setting and function setting use the key operation and characters show in the following figure...

Страница 18: ...e current temperature in furnace operation mode character and setting temperature of overheating prevention device are displayed on respective screens 2 Set the temperature for overheating prevention...

Страница 19: ...ation on the center display screen 3 Start sterilizing operation Press the orange START STOP key for one second The operation starts and the PROGRAM lamp lights on After one hour sterilizing operation...

Страница 20: ...n displays t 1 and the setting temperature screen displays 0 01 Do not change the value Press the ENTER key again The measurement temperature screen displays SV 2 and the setting temperature screen di...

Страница 21: ...setting Also it displays the current setting temperature with blinking The FIXED TEMP lamp blinks too Set the temperature by pressing the 4 Start operation Press the orange START STOP key for about o...

Страница 22: ...ustment is needed 2 Start timer operation Press the START STOP key for one second after deciding the time Timer operation starts with the FIXED TEMP and AUTO STOP lamps lighting on The timer is activa...

Страница 23: ...the ENTER key again The measurement temperature display screen displays the character tim which indicates the timer setting The setting temperature display screen displays the current setting time wi...

Страница 24: ...h the FIXED TEMP and AUTO STOP lamps lighting on Press the START STOP key to terminate the timer operation mode The screen returns to the initial setting screen To correct or check setting Changing th...

Страница 25: ...TO START lamp blinks too Set the temperature using the Press the ENTER key again The measurement temperature display screen displays the character tim which indicates the timer setting The setting tem...

Страница 26: ...heck setting Changing the setting temperature or time during operation is possible by pressing the TIMER key Use the to change the setting value Press the ENTER key respectively after changing the set...

Страница 27: ...apability of increasing or decreasing temperature by 50 within 15 minutes about 30 minutes is needed to increase or decrease temperature by 100 from current temperature Repeat function Repeat function...

Страница 28: ...program mode program pattern Press the PROGRAM key once The measurement temperature display screen displays the previous program mode Press the PROGRAM key again to display the next program mode PrG1...

Страница 29: ...fall capability of unit For example about 75 minutes is needed to increase the temperature from room temperature to 250 for SH401 type 80 minutes for SH601 type When the current temperature is suppos...

Страница 30: ...e set in increments of a minute under 99 hours and 59 minutes It can be set in increment of ten minutes over 100 hours The can change the setting time quickly when it is pressed continuously Press the...

Страница 31: ...mperature screen indicates PS 7 in the example because repeat function is used at the seventh step The step number 1 to 7 can be input in the setting temperature display screen Enter the number 1 in t...

Страница 32: ...n Method Program Operation Programming Preparation Form 1 Please use this form by making copies Register with PrG1 PrG2 PrG3 PAt1 PAt2 PAt3 No Project Name Date Programmer Program Pattern 250 200 150...

Страница 33: ...PAt1 PAt2 PAt3 No Project Name Date Programmer Input Value Temperature Time min Repeat Function Step 1 To Times Step 2 Step 3 Step 4 Step 5 Step 6 Step 7 Step 8 Step 9 Step 10 Step 11 Step 12 Step 13...

Страница 34: ...ature display screen falls by the setting temperature while the temperature on furnace rises When it is set to the minus side the temperature on the measurement temperature display screen rises by the...

Страница 35: ...storm During a thunderstorm turn off the power key immediately then turn off the circuit breaker and the main power If this procedure is not followed fire or electrical shock may be caused When open c...

Страница 36: ...cause trouble Never put sample on the base Fix the shelf on the bracket then set sample on it Recovering after power failure When power is supplied after a power failure the device automatically star...

Страница 37: ...ntenance after returning the temperature of this unit to the normal one Do not disassemble this unit CAUTION Use a well drained soft cloth to wipe dirt on this unit Do not use benzene thinner or clean...

Страница 38: ...equest you to disassemble this unit as possible and recycle the reusable parts considering to the environmental protection The feature components of this unit and materials used are listed below Compo...

Страница 39: ...ort circuit detection ALARM lamp lights on Er 02 appears Triac is in short circuit Make a call for service Heater disconnecting detection ALARM lamp lights on Er 03 appears Heater is disconnected Make...

Страница 40: ...ature fluctuates during the operation Too much samples Wind from air conditioner directly blows The change of ambient temperature is remarkable Samples are too moist The power supply voltage is lower...

Страница 41: ...les office Check following items before contact Model Name of Product Production Number Purchase Date About Trouble in detail as possible Minimum Retention Period of Performance Parts for Repair The m...

Страница 42: ...al functions Lock function Auto recovering after power failure Calibration offset Self diagnostic functions Failure of Sensor heater SSR memory internal communication temperature inputting circuit aut...

Страница 43: ...e Symbol Part name ELB Earth leakage breaker CONT Control board TM Terminal block PIO Display circuit board H1 2 Heater TH1 Sensor for control X Main relay TH2 Sensor for overheating prevention SSR Br...

Страница 44: ...ircuit board VS4 Yamato Scientific 1020000051 Tough card 50mm Yamato Scientific 1130000009 Main relay AHE1254 100V 120V Matsushita 2050000019 SSR TRS5225 Toho Denshi 2160000035 Power cord kit 2 0sq 3p...

Страница 45: ...te Ammonium perchlorate and other perchlorate Potassium peroxide Sodium peroxide Barium peroxide and other inorganic peroxide Potassium nitrate Sodium nitrate Ammonium nitrate and other nitrate Sodium...

Страница 46: ...43...

Страница 47: ......

Страница 48: ...document forbids Unexpected accidents or breakdown may result in Note The contents of this document may be changed in future without notice Any books with missing pages or disorderly binding may be re...

Отзывы: