17
Operation Method
Sterilizing Operation
Change sterilizing
temperature/time
Follow the procedures below to change the sterilizing temperature
or time already registered.
①
Press the PROGRAM key on the initial screen to display the
"PrG1", which indicates the program operation, on the setting
temperature display screen.
②
Press the PROGRAM key again to blink the"PrG1".
③
Press the ENTER key. The measurement temperature screen
displays the "End" and the setting temperature screen displays
"2". Do not change the value.
④
Press the ENTER key again. The measurement temperature
screen displays"SV-1" and the setting temperature screen
displays "180". Change the setting temperature using the "
▲▼
".
⑤
Press the ENTER key again. The measurement temperature
screen displays"t-1" and the setting temperature screen displays
"0.01". Do not change the value.
⑥
Press the ENTER key again. The measurement temperature
screen displays"SV-2" and the setting temperature screen
displays "180". Change the setting temperature same as set in
the step
④
using the "
▲▼
".
⑦
Press the ENTER key again. The measurement temperature
screen displays"t-2" and the setting temperature screen displays
"1.00" (one hour and zero minute). Change the sterilizing time.
⑧
Press the ENTER key to return to the initial screen. This
completes the change of setting.
⑨
Pressing the START key for one second starts operation.
Timer function:
The maximum setting time is "999 hours and 50 minutes".
The time can be set in increments of a minute under 99 hours
and 59 minutes.
It can be set in increment of ten minutes over 100 hours.
The "
▼▲
"can change the setting time quickly when it is pressed
continuously. Press them discontinuously when fine adjustment
is needed.