2
Operation Method
Sequence of Operation
1. Fix the sample containers.
Before turning on the power switch, fix the sample containers firmly.
The quantity of sample containers to be installed is as shown in the
following table.
100
200
300
400
500
1000
Separating Funnel
Rhoto Holder
10
8
8
8
8
6
Centrifugal Tube Holder
Applicable centrifugal tube
Outer dia.: φ16 to φ35× length 110mm to 130mm
Mountable up to 18 pcs. on one side
Test Tube Rack Holder
Dimension of applicable test tube rack
Max. dimension: 238W×121D×105H
Mountable up to 2 racks on one side
2. Turn on the power.
Before turning on the power switch, check if the timer setting dial is
OFF and if the shaking speed adjusting dial is at the lowest speed.
When the power switch is turned on, the power lamp is ON.
3. Set the shaking time.
When the sample container is installed, turn the timer setting dial,
and set it to the required time (min).
In continuous operation, adjust the dial position to "CONT".
4. Adjust the shaking speed.
Set the shaking speed using the shaking speed adjusting dial.
Set the shaking speed referring to the value displayed on the shaking
speed display unit.
CAUTION!
If used at the max shaking speed, such shaking speed could exceed the
standard value in the area of commercial power frequency 60 Hz. If
shaking is performed in the too wide range, or if the main unit vibrates
widely, reduce the speed-regulating dial.
Shaking speed could also rise according to the condition of the load.
In such cases, slightly reduce the speed-regulating dial similarly.
CAUTION!
In the early stage after supplying power, the shaking speed may be
lower than the set value. Perform the running operation for about 15
min. Then, the shaking speed will stabilize.
Capacity (ml)
Holder