8.Troubleshooting Guide
Display and contents
How to take the countermeasure
In case of display of the error as follows in the Error
・
Consumables Replacement Display, take note of
the content of error. Turn OFF the power switch on the right side of main body and close the faucet.
In case of abnormality, parts replacement or system check is necessary. Contact the shop from which
you have purchased the product, our sales office, or our service center.
Note that the serial number must also be informed together with the details of abnormality when
contacting us.
The contact address of the customer service center is shown at the end of this manual.
Alarm
Indication
Causes
Countermeasures
Water quality
error
Stops
all
operations
when
disconnection/short circuit of the
temperature compensation sensor
of the water quality sensor or
temperature
outside
the
measurement range of 0
℃ ~
100
℃
continued beyond the
specified period.
Turn ON the power
switch on the right side
of main body again. If
error persists, call the
service center.
Controller
error
The settings stored in the memory
device in the substrate can not be
read
correctly.
Or
stops
all
operations when an abnormal
condition is detected for the A/D
circuit when the value is abnormal.
Turn ON the power
switch on the right side
of main body again. If
error persists, call the
service center.
Water
leak
error
Stops all operations when water
splashes on the leak sensor.
Turn OFF the power
switch on the right side
of main body, check for
water leak at piping
components and call
the service center.
(See P.32.)
Float
switch
error
Stops all operations when the
sensor input of the float switch is
abnormal.
Turn ON the power
switch on the right side
of main body again. If
error persists, call the
service center.