3. How to use test mode and set the parameters
3-1. How to enter test mode
To enter test mode, remove the sealing screws next to the battery box and press the deep-set switch inside the
small round hole with a pointed nonconductive object. The panel will show decuple-precision weight values.
3-2. How to view information in test mode
In test mode you can confirm the following information.
Press
or
to cycle through the following indications.
Item
Indication
Description
1. TEST
Internal Count
→
Displays a measurement with
precision ten times higher than the
scale interval used in normal
mode, up to a weighing capacity of
30000 to 64000 counts.
2. INIT
Initial Value
→
Displays the internal counts of the
load applied to the platform when
nothing is placed on the weighing
platform. The value changes if the
zero resetting or the zero tracking
is performed.
3. RAW
Raw Count 1
→
Displays a value obtained as a
result of AD conversion.
4. RAW2
Raw Count 2
→
Displays a value obtained by
averaging the values of the Raw
Count 1.
5. ADC0
Option check
→
Displays the status of internal
correction made by the AD
converter.
6
Содержание DP-6900
Страница 29: ...8 Wiring diagram 28 ...