(3) System parameters
System parameter mode is used to change system parameters, which determine the instrument’s
specifications.
◊
How to set system parameters
Operation
Display
When the panel shows Internal Count in test mode, press
and
at the same time.
Then the span adjustment screen appears. Press
and
at the same time. The panel shows the system parameter #b0 with “0”
as its setup value.
Use the following keys when you set system parameters.
: Switches to the next parameter.
: Increases the parameter setting by 1
&
: Return to the previous parameter.
: Decreases the parameter setting by 1.
(IMPORTANT)
When you have changed a setting, be sure to press
to switch to the next parameter before
turning off the power.
Unless you switch to the next parameter, any changes you made to the setting will be lost.
◊
How to set system parameters (Method 2)
Operation
Display
In test mode, select [PArA], and press
.
The parameter setup screen appears.
Use the following keys when you set system parameters.
: Switches between the parameter number and the setup value.
(The flashing field can be edited.)
: Increases the parameter setting by 1
: Decreases the parameter setting by 1.
(IMPORTANT)
When you have changed a setting, be sure to press
or
to switch to the next parameter
before turning off the power.
Unless you switch to the next parameter, any changes you made to the setting will be lost.
12
Содержание DP-6900
Страница 29: ...8 Wiring diagram 28 ...
Страница 33: ...9 2 Exploded view of the weighing platform 32 ...