YCO-1112-04-06
DP-6200
Technical Manual
7
IV Changing the Scale Setup
User Mode
With the scale on, press the
��
and
�
keys to enter the user setup mode. Press the
���
key to exit the user setup mode.
Test Mode
Open the indicator housing by pulling the release tab forward and
swinging the front panel of the housing forward. Short the two test
pins to the right of the CPU on the main board. The Test Mode
ennunciator and internal counts will display. Pressing the
�
key will cause the scale to cycle through the following displays: in-
ternal counts, initial counts, direct raw counts, average raw counts,
battery check/A to D conversion value (should be around 220),
ROM version and a display segment check. Press the
���
key
to exit Test Mode.
System Mode
Enter Test Mode and then press the
��
key and the
�
key.
The display will show system parameter 30. Use the
��
key to
cycle through the system parameters (30 - 40) and then the user
parameters (01 - 09). See the previous page for instructions on
changing the parameters. Press the
��
key and the
�
key
to return to Test Mode or press the
���
key to exit setup.
Factory Mode
Enter Test Mode, press the
��
key and the
���
�����
key to enter
Calibration Mode, and then press the
�
key and the
��
key to enter Factory Mode. The display will show factory parame-
ter 99. Use the
��
key to cycle through the factory parameters
(50 - 99). See the previous page for instructions on changing the
parameters. Press the
��
key and the
�
key to return to
Test Mode or press the
���
key to exit setup.
�
�
Test Mode
System Mode
Factory Mode
Содержание DP-6200
Страница 4: ...YCO 1112 04 06 DP 6200 Technical Manual II Description 2 FIG 1 FIG 2 FIG 3 Name and Function of Parts ...
Страница 21: ...YCO 1112 04 06 DP 6200 Technical Manual VI Wiring Diagram 19 ...
Страница 22: ...YCO 1112 04 06 DP 6200 Technical Manual VIIParts List 20 ...
Страница 23: ...YCO 1112 04 06 DP 6200 Technical Manual 21 VII Parts List ...
Страница 24: ...YCO 1112 04 06 DP 6200 Technical Manual 22 VII Parts List ...