7
DP-6100 GP
Technical Manual
To calibrate scale using
pounds or kilograms, perform the following procedure.
Note: Check and modify Keyword Parameters before attempting calibration.
1)
Enter TEST MODE (refer to page 4).
2)
Wait approximately one minute, then press
CLEAR
.
a)
If "TIME" indicator goes on momentarily* and display reads all zeros >>
(+/- 5), then proceed to step
(3)
(see
Figure 5
). Zero point is set.
b)
If display flashes*
( = "1 more"), then return to step
(2)
.
Note: It may take several minutes to perform
steps
(1)
&
(2)
.
*Often flash is too fast to read or identify.
3)
Place weight # 1 (see
Table 1
) on platform, then press
SETTING
(press only once!).
Important:
Be sure to use correct scale capacity from Table. Scale capacity is marked on indicator.
Display will flash* >>
, then read
(+/- 5) internal counts
(“10000” for 60 lb model). "Span 1" calibration complete.
4)
Place weight # 2 (see
Table 1
) on platform, then press
SETTING
(press only once!).
Display will flash* >>
, then read
(+/- 5) internal counts
(“30000” for the 60 lb model). "Span 2" is complete.
5)
Place weight # 3 (see
Table 1
) on platform, then press
SETTING
(press only once!).
Display will flash* >>
, then read
(+/- 5) internal counts.
"Span 3" is complete.
6)
Remove weight from platform and press
to turn scale off.
7)
Congratulations, calibration is complete! Turn scale on and check accuracy.
TIME
Figure 5
t
h
g
i
e
W
#
)
g
k
0
5
1
(
b
l
0
0
3
)
g
k
0
6
(
b
l
0
5
1
)
g
k
0
3
(
b
l
0
6
1
)
g
k
0
2
(
b
l
0
4
)
g
k
0
1
(
b
l
0
2
)
g
k
5
(
b
l
0
1
2
)
g
k
0
6
(
b
l
0
2
1
)
g
k
0
3
(
b
l
0
6
)
g
k
5
1
(
b
l
0
3
3
)
g
k
0
5
1
(
b
l
0
0
3
)
g
k
5
7
(
b
l
0
5
1
)
g
k
0
3
(
b
l
0
6
Scale Capacity
Table 1
Important:
If calibrating with
lb
, then Keyword 92
must be Parameter Value 001. "
92.001
"
If calibrating with
kg
, then Keyword 92
must be Parameter Value 000. "
92.000
"
Note: Scales with V1.00G CPU
board can only be calibrated with kg.
CALIBRATION