35
TO EDIT SEGMENT
・
There are 12 kinds of setup items of 1 segment.
(
Full option installation
)
No. Name
Setup
range
Option
1 Ramp
time
step
、
1mi
~
999hr
、
end
2
Setup temp
A setup temperature range by the model
3 Soak
time
0
m
i
~
999hr
、
Hold
4 Wait
mode
OFF
、
ON
5 Fan
OFF
、
1
~
10
(
or ON
)※
○
6 Damper
0%
、
25%
、
50%
、
75%
、
100%
○
7 Event
1
OFF
、
ON
○
8 Event
2
OFF
、
ON
○
9 Event
3
OFF
、
ON
○
10 Event
4 OFF
、
ON
○
11
Rep. start segment
No
、
Segment number that 1 - 5 have been
registered
12 Rep
count 1
~
9999
、
Infinity
※
In case “Wind velocity variable device” option is not equipped, you can choose “On” or
“Off” of fan only by “End segment”
・
Concept of segment is shown as below. However “Start temp” corresponds to
“Temperature of the inside of chamber” when you carry out segment number 1 just
after program starts. Other than that case it will be “Setup temp” of segment, which is
carried out most recently.
Lamp time
Sork time
Time
Setup temp
Temp
Start temp
・
Concept of repeat is shown as below .The number of repeat times dose not include the
first practice of repeat section.
Segment No.4
Repeat start
segment :2
Rpeat time:5
Segment No.1
There is no
order for a
repeat.
Segment No.2
There is no
order for a
repeat.
Segment No.3
There is no
order for a
repeat.
Segment No.5
There is no
order for a
repeat.
Repeat section
Repeat×5