![background image](http://html.mh-extra.com/html/yamato/dc801/dc801_instruction-manual_905264020.webp)
18
5. HANDLING PRECAUTIONS
WARNING!
DO NOT operate equipment when abnormalities are detected.
If unit begins emitting smoke or abnormal odors for reasons unknown, turn off main power (ELB)
immediately, disconnect power cable from power supply, and contact Yamato sales office for
assistance. Continuing to operate without addressing abnormalities may cause fire or electric
shock, resulting in serious injury or death. Never attempt to disassemble or repair unit. Repairs
should be always be performed by a certified technician.
DO NOT handle flammable substances without proper ventilation.
Be sure there is adequate ventilation when working with certain flammable substances (such as
ethanol, etc.), which evaporate quickly at or below room temperature, and emit flammable fumes.
Little or no ventilation may cause a fire or explosion resulting in serious injury or death. Refer to
“List of Hazardous Substances” (P.29).
CAUTION!
Cold trap capacity.
Maximum cold trap capacity for DC801 is approximately 1.0L. Exceeding this capacity may
severely degrade performance.
DO NOT climb on equipment.
Do not attempt to climb onto unit or substitute it for a proper step ladder. Units are not designed
to support bodily weight and damage may result. In addition, unit may become unstable and tip
over or fall resulting in equipment damage, serious injury or death.
DO NOT place objects on equipment.
Do not place any objects on unit, which may fall and cause personal injury.
DO NOT operate equipment during thunderstorms.
In the event of a thunderstorm, turn off main power switch (ELB), and disconnect power cable
immediately. A direct lightning strike may cause equipment damage, fire or electric shock,
resulting in serious injury or death.
Overnight and extended storage.
Whenever unit is not in operation, stored overnight or placed in storage, always turn off main
power switch and disconnect power cable. Drain any fluid collected in the cold trap if unit is going
to be out of service for an extended period.
Power supply voltage.
Power supply voltage must be /-5% of the rated voltage.
DO NOT use solvents.
The seals contained in this unit is made from chloroprene rubber and may be damaged by acids,
halogens, aromatics, esters, and oxo solvents. Do not use these compounds with unit.
Starting/restarting refrigeration
When the refrigeration switch is turned from OFF to ON or when refrigeration is resumed by
turning ON the main power switch, a timer delays operation for three minutes to protect the
refrigeration system. Note that this is NOT a failure.