![background image](http://html.mh-extra.com/html/yamato/dc801/dc801_instruction-manual_905264018.webp)
16
4. OPERATION PROCEDURES
Defrost
1. Remove chamber
Remove the flask mounting vacuum chamber from
the trap.
If ice has formed on trap wall, press the defrost
button with the REFRIG switch ON.
2. Remove ice from trap
When ice has melted sufficiently to remove, do so
by pulling up on handle and turn OFF the defrost
switch.
Defrost auto stop activates after approximately 45
minutes and the defrost lamp goes out.
To terminate the defrost function mid-way, turn the
REFRIG switch OFF.
3. Drain water
Remove the drain hose cap, and drain remaining
water from the trap into a drain basin or other
receptacle.
Drain Hose
Handle
Ice
Flask mounting
vacuum chamber
Содержание DC801