Maintenance Method
Daily Inspection and Maintenance
Maintenance of the water bath
Remove foreign substances inside Water Bath as frequency as possible. They may result in circulating
pump malfunction if they are left there.
Must ware pair of gloves at maintenance of this Water Bath.
Replacement of hoses
Replace heat insulation hoses by once every 2(two) years as the standard to keep the
Equipment/Unit in good condition.
Ask Yamato Scientific Co., Ltd. for replacement.
Changing liquid regularly
Change total circulating liquid by every 3 months for city water and by every 6(six) months for Naiburain®
as the standard.
Accumulate fur, alga, and scale in circulating pump so that degrade the Equipment/Unit performance
and then may cause failure, if circulating liquid will degrade.