12
Installation Method
1. Unlock the stopper of the caster.
Pulling up the lever of the stopper for
caster releases the lock.
(Only the two casters in front of the unit are
attached the stopped.)
2. Move the device to the place to be installed.
If there is a step on the floor, the too strong
impact is given to the caster, and could
give the damage. In that case, move the
device by lifting at the step.
3. When the installation place is determined,
pull down the lever of the stopped for
caster, and lock them.
4. Check that the drain cock valve is set at the
“CLOSE” position.
Cover opened status
Close
Drain cock
Overflow
Open
Drain cock
valve
REFRIGE
Temperature c
ontroller
Caster
Stoppe
r
Draining position
Operating