1 - 7
System Configuration for Analog Output (4 - 20 mA DC output)
continued
System configura-
tion WITH the
communication
function
Figure 1-4 shows a sample system configuration in which the instantaneous
flow rate measured by the unit is output with a 4 - 20 mA DC analog signal.
In order to enable communications, a DC power supply and a resistance of
250
Ω
or more must be installed on the receiving side.
Figure 1-4 System configuration for analog output WITH the communica-
tion function (example setup)
Host
control
system
(1 - 5 V
input)
S-SFC
AC or DC
power
supply
24 V DC power
supply
4 - 20 mA
analog output
250
Ω
resis-
tance
Smart
communicator
• Smart Electromagnetic Flowmeter (device): Measures flow rate and out-
puts an analog signal instantaneous flow rate.
• Smart Communicator (S-SFC): Communicates with the device to read data
and change the device settings.
Device