Dental X-Ray Unit
Z70
70kVp - 7mA
10
Operation Manual
Items to be checked:
- Check if all product labels are proper in place
- Check if the Tubehead is free from oil residues
- Check if coiled cable and X-Ray button are not broken and working properly
- Check for external damage in the Tubehead cover
- Check for the arm balancing (Up and Down)
- Check (in case wall model) if the screws from the wall support and or wall plate are
proper installed
- Check (in case wall model) if the vertical alignment of the wall plate/support is on level
- Check (in case mobile model) if all base screw are well fixed
- Check if all equipment rotate limit are working properly
The checking should be performed before any operation session. In case of irregularities,
contact immediatly the Technical Service.
3.8. Equipment Test
After the product installation (technicians only), test all product functions, following all steps
described in this Manual. If there are some doubt, enter immediatly in contact with the
Technical Service Support
3.9. Equipment Transportation and Storage Condition
- Environmental temperatura: - 20
°C (- 4° F) ~ + 55° C (131° F)
- Relative humidity:10% ~ 80%
- Atmospheric pressure: 500HPa ~ 1060HPa
- The equipment should be stored in a dry, airy and free of impurities
- Do not expose the equipment to strong vibrations
Warning
3.10. Equipment Use
- Allow enough space for the product use and operation
- Do not hang objects on the equipment
- For legal effects, the useful life of the product is 5 years from date of manufacture